Universal chiller iChiller-Product features of IC036
LCD display, touch button operation, user-friendly operation design.
High performance circulating water pump, dual water purification design of inlet filtration and pre outlet filtration circulating water system, prevents impurities from entering the circulating pump, prolongs the service life of the pump, and also ensures clean water quality entering the external main equipment;
Pre installed water outlet filter, standard tools are standard, allowing users to easily detect filter element dirt and promptly address itreplaceFilter element, easy to replace filter element;
Equipped with an adjustable pressure limiting pump protection valve as standard, when the system pressure does not exceed the pressure limiting value, the pump flow is fully supplied to the external main equipment. When the external equipment is disconnected and does not require circulating water, the system pressure increases. When it exceeds the pressure limiting value, the pressure limiting valve opens and the pump flow returns to the water tank from the pressure limiting valve, protecting the circulating pump and extending its service life.
System stability and reliability: High performance and high-quality spare parts with stable quality and reliable performance.
Multiple safety protections: temperature protection, water pressure protection, overheating and overcurrent protection. When an alarm is triggered, both light and sound can be used to remind.
The refrigeration series adopts imported compressors and silent fans, which make the whole machine work with low noise and suitable for laboratory environment applications.
Optional flow switch alarm signal output, timely alarm when the circulation pump fails and stops or the pipeline leaks, causing insufficient peripheral circulation flow, to protect the safety of external main equipment;
Optional hardware temperature protection, when the refrigeration system fails and causes a temperature rise, outputs an alarm signal to external devices through the set hardware protection switch to protect the safety of external devices.
Universal chiller iChiller-IC036'sapplication
Application of supporting instruments, small industrial equipment, and lasers in the field of life sciences.
Supporting analytical instruments: supporting atomic absorption, ICP, Electron microscopy, Kjeldahl nitrogen determination, Soxhlet extraction, ion thinning analyzer, molecular pump, diffusion pump, annealing furnace, small distillation apparatus, reaction kettle, fermentation tank, etc.
IC036Main technical parameters
| Product model | [-] | IC036 |
| Working temperature range | [°C] | 5...35 |
| temperature stability | [±K] | ±1.0 |
| Temperature Display | [-] | LCD |
| display precision | [°C] | 0.1 |
| Cooling capacity (25 ° C) | [W] | 3600 |
| pump pressure | [bar] | 2.7 |
| pump flow rate | [L/min] | 75 |
| Pump connection interface | [mm] | RC1/2 (4-point internal teeth) interface, compatible with various specifications of pipelines |
| tank capacity | [L] | 25 |
| Sink size | [mm] | φ260*470 |
| Sink opening | [mm] | φ210*470 |
| Dimensions (WxDxH) | [mm] | 530x680x820 |
| Instrument fixing method | [-] | floor-standing |
| Allow continuous operation | [%] | 100 |
| Allow ambient temperature | [°C] | 5..32 |
| Allow relative humidity | [%] | 80 |
Power supply voltage/frequency [VHZ] 200...230VAC/50Hz