Product parameters Brand: Fubus FURBS Model: FK-1200K Weight: 8kg Measurement range: unlimited
Large scale solid density tester (customizable)
FK-1200F

Suitable for industries such as rubber, plastics, hard alloys, ceramics, powder metallurgy, casting parts, aluminum foundries, and precious metal hardware recycling.
Principle: This machine is composed of American precision scales combined with professional specific gravity accessories; According to ASTM D297-93,
D792-00, D618, D891, GB/T1033, JISK6530, ISO2781 standards. Using the Archimedes principle buoyancy method,
Accurate and direct reading of measurement values.
|
model:
|
FK-1200F
|
FK-2000K
|
|
Weighing accuracy:
|
0.01g
|
0.01g
|
|
Weighing range:
|
0.01g~ 1200g
|
0.01g~ 2000g
|
|
Specific gravity accuracy:
|
0.001 g/cm3
|
0.001 g/cm3
|
|
Measurement time:
|
About 10 seconds
|
About 10 seconds
|
|
set up:
|
Temperature compensation setting, solution compensation setting
|
Temperature compensation setting, solution compensation setting
|
1. The relative relationship between the compressive or tensile strength of a product and its overall density. Large parts cannot be segmented for testing, and the influence of the cutting surface during segmentation testing can cause inaccurate density measurement
2. FK-2000F/3000F is a specific gravity meter designed for measuring large parts, which is different from the general hanging measurement method. It is designed to use a large specific gravity measuring platform.
3. The space of the internal sink reaches 22 * 18 * 14cm.
4. The maximum weighing capacity can reach 5000g/10000g, and the density analysis is 0.001g/cm3
5. Testing density can indicate changes in material composition or defects in the product, such as cracks or bubbles
6. The melting quality in aluminum casting plants is monitored through two environmental samples: one is under air pressure
Another one is under a pressure of 80 mbar. Once they are set and cooled, the density of the sample is determined, and the ratio of the two density values is an important basis for providing melting purity.

Step 1: Place the tested sample on the upper tray, and press the Mode key to automatically remember the weight in the air W1 (machine automatically records)
Step 2: Place the test substance into the water tray and press the Mode key to automatically remember the weight of the water. After about 3 seconds, the density (specific gravity) value will be displayed directly