Laboratory liquid particle counter, water particle detector

NegotiableUpdate on 05/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The laboratory liquid particle counter and water particle detector can provide fast, accurate, reliable, and reproducible detection results, suitable for detecting the size and quantity of solid particles in various low viscosity liquids. Widely used in the fields of aerospace, power semiconductors, petrochemicals, automotive manufacturing, etc. for low viscosity oil contamination analysis, precision component cleanliness evaluation, chemical reagent and cleaning agent particle detection, as well as purified liquid and ultrapure water particle determination.

Product Details

Laboratory liquid particle counter, water particle detector

This instrument is developed based on the principle of light blocking counting. It can provide fast, accurate, reliable, and reproducible detection results, suitable for detecting the size and quantity of solid particles in various low viscosity liquids. Widely used in the fields of aerospace, power semiconductors, petrochemicals, automotive manufacturing, etc. for low viscosity oil contamination analysis, precision component cleanliness evaluation, chemical reagent and cleaning agent particle detection, as well as purified liquid and ultrapure water particle determination.

Laboratory liquid particle counter, water particle detectorMain Features

High precision laser sensor with a wide testing range, stable performance, low noise, and high resolution

Adopting high-pressure injection pump sampling method, the sampling volume can be set by oneself, the injection speed is stable, and the sampling accuracy is high

Built in multiple calibration curves, compatible with all commonly used domestic and international standards for calibration

L supports custom standard testing, allowing for naming and grading of custom standards

According to customer requirements, up to 64 detection channels can be set up simultaneously

Built in data analysis system, one test can provide all standard test data and pollution levels

L can perform cleaning detection, pre-set target cleaning requirements particle concentration, display cleaning status in real-time during the cleaning process, and automatically stop when the requirements are met

L lifting sensor design, suitable for various sampling containers, meets the detection requirements of different industries

L can perform manual, automatic, and semi counting calibration

Windows operating interface, built-in database, unlimited data storage, can be searched and queried using various methods such as testing standards, sample names, sample batch numbers, testing personnel, testing time, etc

Multi level account permission management, allowing for separate account permissions and supporting digital signatures

A control system supports multiple sensor settings, and the system control software can be installed on a tablet, laptop, or desktop computer

Support scanning gun information input, test results can be printed using micro or desktop printers, can be exported in PDF, Excel, and other formats, and can be transmitted through USB, WIFI, and other methods

L has audit tracking function

L has a particle size distribution curve, which facilitates particle size analysis

technical indicators

Light source: semiconductor laser

Particle size range: 0.8 μ m to 600 μ m

L detection channels: 8-64 channels are optional (default 64 at the factory), and the particle size can be set arbitrarily

Sensitivity: 0.8 μ m (ISO4402) or 3 μ m (c) (GB/T18854, ISO11171)

Resolution: better than 6%

Repeatability: RSD<2%

Sample detection viscosity: ≤ 100cSt

Sampling volume: 0.2-6000mL, with an interval of 0.1mL

Sampling accuracy: better than ± 0.2%

Sampling speed: 5-80mL/min

Limit of coincidence error: 40000 particles/mL

Test sample temperature: 0 ℃~80 ℃

Working temperature: -20 ℃~60 ℃

Storage temperature: -30 ℃~80 ℃

Power supply: AC 90-240V, 50/60Hz

Host size: 310mm (L) × 340mm (W) × 440mm (H)

Net weight of the host: ≯ 12kg