- Phone
-
Address
Yizhuang Economic and Technological Development Zone, Beijing
Beijing Ruiyuan Lixin Technology Co., Ltd
Yizhuang Economic and Technological Development Zone, Beijing
Our company has a laboratory instrument moving team that specializes in handling precision and weight instruments in the laboratory. Especially without cargo elevators and equipment that requires manual handling. For example, biosafety cabinets, shaking tables, carbon dioxide incubators, liquid nitrogen tanks, -80 ℃ refrigerators, disassembly and assembly of various laboratory furniture, laboratory benches, fume hoods, etc.
The laboratory relocation requires a systematic planning and full process risk management, with a focus on the three core links of hazardous material disposal, equipment protection, and qualification maintenance1
Pre relocation planning and management
Establish a relocation team consisting of equipment engineers, safety specialists, and project leaders, clarify division of labor, and file a list of equipment. 23
:
Measure the water and electricity load, ventilation system, and gas cylinder layout of the new laboratory to ensure compliance with GB/T 32146 "Technical Requirements for Laboratory Design and Construction". 4
Plan the equipment flow line, verify the load-bearing capacity of the freight elevator and the size of the passage (it is recommended to leave a 20cm safety margin). 25
:
Triple sealed packaging (inner glass bottle+middle cushioning material+outer explosion-proof box) is used for hazardous chemicals, and a separate leak emergency kit is provided for highly corrosive liquids. 6
Radioactive materials are entrusted to CNAS qualified institutions for processing, and biological samples are transported under dual temperature monitoring. 78
Risk management during relocation
:
Class I equipment such as mass spectrometers use air cushion shock absorbers. Before disassembly, mark the position of optical components and take photos for archiving. 78
Seal the interface of the liquid chromatography pipeline with cling film to prevent foreign objects from entering. 1
:
A dedicated transport vehicle equipped with a temperature and humidity recorder and GPS positioning, with a speed controlled within 40km/h. 69
Vehicles transporting hazardous chemicals must hold a "Road Dangerous Goods Transport License" and the driver must have the qualifications for transporting hazardous chemicals. 3
Commissioning and acceptance after relocation
:
The electronic balance needs to be preheated for 24 hours before calibration, and the liquid chromatograph uses standard samples for retention time drift testing (allowable deviation ≤ 2%). 48
Biosafety cabinets are used for detecting airflow patterns and descending wind speeds. 10
:
Submit the "Address Change Application Form" to CNAS within 20 working days, and stop using the approved logo until it passes the on-site review. 13
Keep performance verification records before and after equipment relocation for future reference, with a period of no less than 3 years.