KV100DautomaticFoldable tube typeKinematic viscosity meterComposed of two independent KV100 measurement units, they can be set to different temperatures for synchronous testing. Each measuring unit adopts a single channel standard folded tube glass capillary viscometer,passDistributed HeatingprecisionMetal bath achieves viscosity controlplanFast constant temperature,instrumentIt integrates high-precision temperature control system, high-precision measurement system, and efficient automatic cleaning system, mainly used for rapid laboratory testing and instrumentationAccurate dataFully functional and easy to operate,The viscometer can be quickly replaced to cover different viscosity ranges,It can quickly and directly measure the kinematic viscosity of various new oils, in use oils, and other transparent and opaque Newtonian liquids at multiple temperatures.

KV100DautomaticFoldable tube typeKinematic viscosity meterperformance metrics
According to the standards: ASTM D7279, D445, D446, NB/SH/T 0956-2017, T/CEC 127-2016, the measurement results meet and exceed the requirements of GB/T265 standard.
Measurement range: 0.3-6000mm2/s, with a single viscosity tube measuring a span of 10 times.
Sample dosage: 0.3-1 mL
Temperature control range:40~100℃
Temperature control accuracy: ± 002℃
Timing accuracy: ± 0.01S
Repeatability:≤0.68%
Whole machine power:≤400W
Host size:480210400( mm)(1)
working environment
Working power supply:AC220V±10% 50Hz±10%
Environmental temperature: 10-28 ℃
Relative humidity:<80% RH
Other: There is no strong vibration, airflow, strong electromagnetic interference, or corrosive gas around the laboratory.
Instrument Features
The instrument consists of two independent measuring units, each of which can perform viscosity measurements at any temperature point between 40-100 ℃.
Each measuring unit adoptsdistributeddirect currentheatedprecisionMetal bath,No need for constant temperature liquid, small volume, constant temperatureFast speed. warm upThe top speed can reach 8℃/min,The rapid cooling speed can reach5 ℃/min.
Each measuring unitadopttwoPT1000High precision temperature sensorRealize precise zone constant temperature and uniform temperature field.
The entire system adopts a DC 24V working voltage, which is safe and reliable.
Equipped with video monitoring function and optional micro camera for real-time monitoring.
Rich alarm and protection functions,Dual over temperature protection and over temperature alarm for both software and hardware、Bubble alarm, pollution alarm, blockage alarm.
Fast measurement speed.Non standard mode can be up to 15The results are usually obtained within 3 minutes, and the entire process of injection, measurement, cleaning, drying, and result calculation generally takes no more than 3 minutes7minute
8High degree of automation. Measurement, cleaning, drying, and result calculation are all automatically completed.
9There are many types of samples for measurement. Can measure transparent and opaque samples, including gasoline, diesel, keroseneTurbine oilCutting fluid, heat transfer oil, additives, lubricating oil, new oil and in use oil and other Newtonian liquids.
10Fully automatic cleaning, fast cleaning, and low cleaning cost. A cleaning process generally consumes no more than 10 milliliters of cleaning solution.
11Multiple auxiliary functions. Includes: constant calibration, temperature calibration, internal clock timing verification, automatic calculation of viscosity index.
12ViscometerEasy to replace, canQuick replacement.
13、It has eleven self diagnostic functions.
14. Can store eight sets of viscometer constants.