Hitech Instruments Co., LtdK522 Hydrogen Purity AnalyzerUse thermal conductivity sensors to accurately detect changes in a specific component in binary or quasi binary gas mixtures (i.e., where only one component changes in the mixture). itsThe sensor has a high sensitivity, low heat capacity, and non consumable component that requires no maintenance and has a long lifespan.
K522 Hydrogen Purity AnalyzerIt is a 'no display' instrument. The standard configuration of K522 consists of electronic modules, sensors, sample chambers, and increased safety junction boxes. This instrument provides 4-20mA output and RS232 communication.
It can be accessed through softwareK522 OEM analyzerVarious configuration parameters such as detection gas, background gas, detection range, analog output, etc. can be set, and real-time readings of the process can also be verified and recorded through software.
It can be programmed to detect 0-5%, 0-10%, 0-30%, 0-100%, 95-100% of H2 or He in N2, Ar, or O2 background gases, as well as other gases and ranges. Please feel free to inquire.
K522 Hydrogen Purity Analyzertechnical parameters
Environmental temperature range
-20°C ~ +50°C
Requirements for sample gas
The standard configuration of this instrument does not include a sampling system and a sampling chamber. The sample gas must not agglomerate and will not cause corrosion when passing parallel to the sensor sintering surface (i.e. not directly facing the sintering surface).
Sensor: Stainless steel thermal conductivity meter sealed with air reference gas
Gas detection and its range: 0-5% hydrogen in nitrogen or argon or hydrogen in oxygen, 0-10% hydrogen in nitrogen or argon or hydrogen in oxygen, 0-30% hydrogen in nitrogen or argon or hydrogen in oxygen, 0-100% hydrogen in nitrogen or argon or hydrogen in oxygen, 95-100% hydrogen in nitrogen or argon or hydrogen in oxygen, 0-100% helium in nitrogen or helium in argon or hydrogen in oxygen, 75-100% nitrogen or hydrogen in oxygen, and 0-100% nitrogen in nitrous oxide.Other ranges and gases can also be used. Please consult customer service.
Sample gas pressure: Working pressure: 1 atmosphere+/-100mbar, maximum absolute pressure range: 0-10bar.
Sample gas flow rate: When configuring the sample gas chamber provided by Hitech, the recommended flow rate is 100 ml/min -300 ml/min.
user interface
RS232 interface: special rate: 9600, with hardware flow control (CTS/RTS). ASCII protocol. Randomly provide PC terminal software for controlling and configuring K522 instrument, part number 290-7022.
MA output: 4-20 mA analog output corresponding to the gas range, which can be selected through the interface software using simple ASCII commands.
Electrical specifications
Power supply: 24V DC+/-10%, maximum power consumption of 10W.
Analog output: * Isolated internal power supply 4-20 mA current source. 1KV isolation voltage. The minimum load is 0 Ω and the maximum load is 300 Ω.
Connecting cable: The maximum cable length between the electronic component and the sensor is 30 meters. The maximum impedance is 0.5 Ω.
Accuracy and Resolution:
gas detection
Accuracy: ± 2% of the set range
Reproducibility: ± 0.5% of the set range
Resolution: 0.1% of the set range
analog output
Resolution of the entire set range: 0.01 mA
Accuracy:+/-0.1% of the set range
Linearity:+/-0.1%
Minimum analog output below range: 3.8 mA
Maximum analog output beyond range: 20.5 mA
Upper error range: 21.0-24.0 mA