Jinan 10T liquid chromatography instrument
Jinan 10T liquid chromatography instrumentProduct Introduction
The new LC-10T liquid chromatograph consists of an LC-10Tvp infusion pump, SPD-10Tvp UV variable wavelength detector, an injector (manual or automatic), a chromatographic column, and a chromatographic data workstation. It is recommended to use Survey software for the chromatographic data workstation. It adopts a digital communication system, wide dynamic range, and can perform reverse control of liquid chromatography on a computer. Can be applied in analytical fields such as medicine, food, chemical engineering, and environment. The new version of the liquid chromatograph complies with GLP/GMP and 21CFR Part 11 specifications
Introduction to Standard Configuration Product Characteristics, Indicators, and Parameters Software
LC-10Tvp liquid phase series 2016 appearance, equipped with a new version of chromatography workstation survey that meets GMP requirements, electronic signature, user permission allocation, audit tracking, system adaptability and other functions.
Appearance and craftsmanship:
Changes in door opening design
Reasonable pipeline design
Cooling system
Change tray design
Improved conductive silicone button design
Change the instrument signal communication method
| LC-10Tvp high-pressure constant current infusion pump | 1 unit |
| SPD-10Tvp UV detector | 1 unit |
| Solvent tray | 1 unit |
| 8725i manual injection valve | 1 piece |
| Chromatography workstation | 1 set |
| Microsampler | 1 tube (50 μ L) |
| Liquid chromatography column | 1 piece (C18 4.6 * 250mm, 5 μ m) |
| Injection bracket | 1 piece |
LC-10Tvp high-pressure constant current infusion pump:
| Infusion method | Micro volume series double plunger |
| Infusion pressure | 6000psi |
| Flow setting range | 0.001~9.999ml/min (adjust flow rate in 0.001ml/min steps) |
| Error in flow setting value | ≤0.5% |
| Flow stability error | ≤0.2%RSD |
| pressure fluctuation | ± 1% (0-100%, 2-liquid gradient of water/acetone solution) |
| Pump sealing performance | Less than 0.1MPa, flow rate 1mL/min, pressure drop less than 5MPa |
| Time program function | have |
| size | W260*H130*D420mm |
SPD-10Tvp UV Visible Variable Wavelength Detector:
| Wavelength Range | 190nm~700nm |
| Wavelength indication error | ≤±1nm |
| Wavelength repeatability error | ≤±0.1nm |
| Dynamic noise | ≤± 0.75 × 10-5AU (methanol, 1mL/min, 254nm, 20 ℃) |
| Static noise | ≤± 0.5 × 10-5AU (empty pool, response time 1 second, 20 ℃) |
| Dynamic baseline drift | ≤± 1 × 10-4AU/h (methanol, 1mL/min, 254nm, 20 ℃) |
| Static baseline drift | ≤ 0.5 × 10-4AU/h (empty pool, response time 1 second, 20 ℃) |
| Linear range | ≥104 |
| Detect concentration | ≤4×10-9G/mL (naphthalene/methanol solution) |
| Qualitative repeatability | RSD ≤0.1% |
| Quantitative repeatability | RSD ≤0.5% |
| Spectral bandwidth | 6nm |
| Flow pool volume | 10μL |
| Optical path length | 10mm |
| Time program function | have |
| size | W260*H130*D420mm |