The Jepp Jensprima water quality total hardness tester PACON 4200 adopts the principle of titration colorimetry, with a compact structure, easy operation, and accurate measurement. It is an entry-level choice for water softening systems and boiler room water quality monitoring. $r $n $r $n Measurement parameters: Total hardness $r $n Typical applications: boiler water supply and boiler water, circulating water, process water

Product Details

Jensprima water quality total hardness tester PACON 4200Features:

Customized boiler room design with high cost-effectiveness

Automatic measurement and automatic cleaning

Graphic backlit LCD display for external signal input control measurement, used for starting or stopping analysis externally

4-20mA output, 2-channel relay output

Display unit ppm CaCO3 continuous measurement or interval measurement (5-30 minutes)

SD card data storage (historical data, fault records)

Jensprima water quality total hardness tester PACON 4200Specifications and Technical Parameters:

Measurement range: 0.2-500.0 ppm CaCO3 (see reagent type)

Measurement time: about 3 minutes, depending on the hardness of the water and the set flushing time

Analysis cycle: continuous measurement/interval measurement (5-30 minutes)/external start signal

Rinse time: 120 seconds

Water consumption: approximately 1000ml/analysis

Display: Backlit LCD display graphics, numerical unit: ppm CaCO3

Current output: 4-20mA, Max. 750 Ω

Relay output: 2-channel passive relay output NC, 250VAC 4A

Input: External switch signal start analysis Basic parameters: Measurement principle: Titration colorimetric method

Environmental temperature: 5-45 ℃ Water sample temperature: 5-40 ℃

Water sample pressure: 0.5-5 bar, recommended 1-2 bar. If it exceeds 2 bar, it is recommended to install a pressure reducing valve

Water quality requirements: colorless, no suspended solids, no bubbles, pH 4-10.5, iron:<3ppm, copper:<0.2ppm, aluminum:<0.1ppm, manganese:<0.2ppm

Inlet/outlet connection: 6mm outer diameter hose

Humidity: 20-90% RH, indoor installation

Power supply: 24VDC, 25VA (during operation)

Size/Weight: 250 x 360 x 110mm, ca. 2.5Kg (including outer box shell)