Japan Sanfeng roundness tester cylindrical measuring instrument RA-1600

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The compact and wide measurement range of the Japanese Sanfeng roundness measuring instrument RA-1600 cylindrical measuring instrument ensures that its accuracy can be comparable to higher-level models. Maximum measured diameter:; 280mm vertical movement distance: 300mm maximum worktable load: 25kg radial rotation accuracy: (0.02+6H/10000) μ m axial rotation accuracy: (0.02+6X/10000) μ m

Product Details

Japan Sanfeng roundness tester cylindrical measuring instrument RA-1600

1. The compact size and wide measurement range ensure that its accuracy is comparable to higher-level models. Maximum measured diameter: ø 280mm Vertical movement distance: 300mm Maximum worktable load: 25kg Radial rotation accuracy: (0.02+6H/10000) μ m Axial rotation accuracy: (0.02+6X/10000) μ m
2. The D.A.T. (digital adjustment table) function helps with manual workpiece centering and leveling.
3. The detection section provides safety devices as standard features. The detector is equipped with collision sensing function (in the vertical direction) to prevent destructive collisions on the Z-axis. In addition, it is equipped with an over limit prevention function, which will stop the system when the displacement of the detector exceeds its range.
4. A sliding mechanism can be installed inside the detector bracket (optional sliding detector bracket). It is difficult to achieve standard detectors that can measure deep hole thick walled workpieces with a single click.
5. If roundness/cylindricity analysis software (ROUNDPAK) is used, the measurement results can be displayed graphically.

Japan Sanfeng roundness tester cylindrical measuring instrument RA-1600Performance Introduction:


Small size and wide measurement range ensure its accuracyComparable to higher-level models.Maximum measured diameter: ø 280mmVertical movement distance: 300mmMaximum worktable load: 25kgRadial rotation accuracy: (0.02+6H/10000) μ mAxial rotation accuracy: (0.02+6X/10000) μ m


The D.A.T. (digital adjustment table) function helps with manual workpiece adjustmentHeart and tone level.


The detection section provides safety devices as standard features. inspectThe output device is equipped with collision sensing function (in the vertical direction)Prevent destructive collisions on the Z-axis. In addition, there are alsoEquipped with overload prevention function,When the displacement of the detector exceeds its limitWhen the range is reached, the system will be stoppedThe operation.


Can be installed inside the detector bracketSliding mechanism (optional sliding inspection)Dispenser bracket). Capable of measuring deep hole thickness with a single clickIt is difficult to implement standard detectors.


If roundness/cylindricity is usedAnalysis software (ROUNDPAK),Can display measurements graphicallyQuantitative results.


Product Specifications
model RA-1600






Rotating worktable
Rotation accuracy
(JIS B 7451-1997)
radial (0.02+6H/10000) µ m H: for measuring height (mm)
axial (0.02+6X/10000) µ m X: Distance to the axis of the rotating worktable (mm)
rotation speed 4, 6, 10rpm
Effective diameter of workbench ø150mm
Range of heart/level adjustment D.A.T.
Heart regulating range ±3mm
Adjust the level range ±1º
Maximum load-bearing capacity of countertop 25kg
Maximum measured diameter ø280mm
Maximum workpiece diameter ø560mm



Vertical axis movement
Drive straightness (λ c2.5) Narrow range: 0.20 µ m/100mm, wide range: 0.30/300mm
Parallelism with the turntable axis (reference to the busbar element) 1.5µm/300mm
positioning speed Maximum: 15mm/s (measured at 0.5, 1.0, 2.0, 5.0 mm/s)
Vertical travel 300mm
Maximum detection height 300mm*1
Maximum detection depth (when using a standard measuring rod) ø 32 × 91mm (depth), ø 7 × 50mm (depth)


X-axis
Straightness (λ c2.5) 2.7µm/140mm
Verticality to the turntable axis (reference to the busbar element) 1.6 µm/140mm
range of movement 165mm (-25mm -+140mm turntable center)
movement speed Maximum: 8mm/s (measured at 0.5, 1.0, 2.0, 5.0 mm/s)



detector
force measurement 10 - 50mN, Levels 1-5
Standard measuring rod, material ø 1.6mm hard alloy
scope standard ±400µm/±40µm/±4µm
track ±5mm
other Simple touch switch for measuring inner/outer diameter, Z-axis collision detection function, and angle scale mark on the measuring head (± 45 º)



other
power supply 100-240VAC
power consumption 80W (excluding PC and printer)
atmospheric pressure 0.39MPa
air consumption 22L/min standard state (gas source above 80L/min)
host weight 170kg
*1: If the measurement height of the workpiece is less than 20mm, an auxiliary workbench option is required.