Intelligent small flow particulate matter sampler for ambient air 16.67

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

This instrument is used to collect samples of inhalable particulate matter (PM10) or fine particulate matter (PM2.5) in the ambient air. It can be used for routine aerosol monitoring by departments such as environmental protection, health, labor, safety supervision, military, scientific research, and education. Intelligent small flow particulate matter sampler for ambient air 16.67

Product Details

1、 Execution standardsIntelligent small flow particulate matter sampler for ambient air 16.67

HJ93-2013 Environmental Air Particulate MatterTechnical Requirements and Testing Methods for PM10 and PM2.5 Samplers

HJ/T 374-2007 "Technical Requirements and Testing Methods for Total Suspended Particle Sampling Instruments"

HJ618-2011 "Determination of PM10 and PM2.5 in ambient air by weight method"

HJ656-2013 Technical Specification for Manual Monitoring Method (Weight Method) of Environmental Air Particulate Matter (PM2.5)

JJG 943-2011 "Total Suspended Particle Sampler"

2、 Purpose:Intelligent small flow particulate matter sampler for ambient air 16.67

This instrument is used to collect inhalable particulate matter in the ambient airInstruments for samples of PM10 or fine particulate matter (PM2.5). It can be used for routine aerosol monitoring by departments such as environmental protection, health, labor, safety supervision, military, scientific research, and education.

3、 Performance characteristics

Unique system design for collecting inhalable particulate matter in the atmosphere(PM10) or fine particulate matter (PM2.5) samples

Automatic control of sampling flow rate: using electronic flowmeter, microcomputer system detects sampling flow rate, and automatically compensates for flow rate changes caused by voltage fluctuations, resistance, and temperature changes

Automatically calculate the cumulative sampling volume and convert the standard sampling volume based on air pressure and temperature

During the sampling process, power outages automatically save the number of jobs According to reports, sampling can be restored after a call

Precision core pump, corrosion-resistant, ultra-low noise, maintenance free continuous operation, adaptable to various working conditions, with overload protection function

High quality dust filter screen, achieving efficient filtration function, protecting the sampling pump and gas path

AC/DC dual-use, built-in lithium battery

Can achieve immediate sampling, timed sampling, and equidistant sampling

Wide temperature and high brightness display screen, suitable for high-altitude regions, with a popular software display interface, achieving good human-computer interaction

Sampling data can be stored and printed, making it easy to grasp real-time data

The deviation between the sampling flow rate of the instrument and the set flow rate exceeds± 10%, and if the duration exceeds 60 seconds, the instrument will automatically stop sampling and record this condition

Can provide users with the current date and time at any time for easy operation

Intelligent software calibration function

The sampling head is made of aluminum alloy material and has anti-static adsorption

Modular design, reliable rain and snow protection function

4、 Technical parameters

sampling flow:16.67 L/min, Resolution 0.01 L/min, accuracy not exceeding ± 2%

Sampling time:99 hours and 99 minutes can be set arbitrarily, with a resolution of 1 minute and a timing error of no more than ± 1 second within 20 minutes

Temperature before flowmeter:(-55~125) ℃, resolution of 0.1 ℃, accuracy not exceeding ± 2.5 ℃

Traffic repeatability: not exceeding2%

Atmospheric pressure:(50~130)kPa, Resolution of 0.01kPa, accuracy not exceeding ± 500Pa

Instrument noise:≤60dB(A)

Load capacity: Overcoming resistance≤12kPa

Dimensions: LengthX Width x Height 210mm x 270mm x 270mm

Battery charger:DC29.4V 2A

Battery cycle life:≥ 1000 times

Working power supply: built-in lithium battery

Host weight: approximately7.5kg

Power consumption:<40W

Sampling head indicators

PM2.5 cutting characteristics: Da50 = (2.5 ± 0.2) μm, δg = (1.2 ± 0.1) μm

PM10 cutting characteristics: Da50 = (10 ± 0.5) μm, δg = (1.5 ± 0.1) μm

Entrance speed:0.3 m/s

Cutting point flow rate:16.67 L/min

Filter membrane diameter:Φ 47 mm (effective diameter Φ 40 mm)