Welcome Customer !

Membership

Help

Nanjing Beiao Instrument Manufacturing Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Integrated synchronous thermal analyzer

NegotiableUpdate on 01/12
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The comprehensive synchronous thermal analyzer combines thermal analysis (TG) with differential thermal analysis (DTA) or differential scanning calorimetry (DSC), and can simultaneously obtain information on TG and DTA or DSC using the same sample in the same measurement. The STA-6000 comprehensive synchronous thermal analyzer mainly measures physical and chemical changes related to heat, such as the melting point, melting heat, crystallization and crystallization heat, phase change reaction heat, thermal stability (oxidation induction period), glass transition temperature, adsorption and desorption, component content analysis, decomposition, combination, dehydration, additives and other changes of substances for research.
Product Details
STA-6000 comprehensive synchronous thermal analyzerProduct Introduction:
Integrated thermal analysis combines thermal analysis (TG) with differential thermal analysis (DTA) or differential scanning calorimetry (DSC). By using the same sample in the same measurement, TG and DTA or DSC information can be obtained synchronously. Comprehensive thermal analysis mainly measures physical and chemical changes related to heat, such as the melting point, melting heat, crystallization and crystallization heat, phase change reaction heat, thermal stability (oxidation induction period), glass transition temperature, adsorption and desorption, composition analysis, decomposition, synthesis, dehydration, additives and other changes of substances for research. Flexible and diverse designs combined with a wide range of options are the ideal tools in your laboratory. Instruments are widely used in most material fields, including plastics, rubber, synthetic resins, fibers, coatings, grease ceramics, cement, glass, refractory materials, fuels, pharmaceuticals, food, refractory materials, etc.

综合同步热分析仪

meet the standard
GB/T 14837-2018 Rubber and rubber products - Determination of composition of vulcanized and uncured rubber by thermogravimetric analysis (Parts 1-3)
JB/T 6856-1993 Thermogravimetric Differential Thermal Analyzer
GB/T 36246-2018 "2018 Synthetic Material Surface Sports Field for Primary and Secondary Schools"
YB/T 4328 Test Method for Free Calcium Oxide in Steel Slag
GB/T 2951.41-2008 General Test Methods for Insulation and Sheath Materials of Cables and Optical Cables Part 41: Polyethylene and
Special test method for polypropylene mixtures - Environmental stress cracking resistance test.


STA-6000 comprehensive synchronous thermal analyzerTechnical Specifications:

temperature range Room temperature~1150 ℃
temperature resolution 0.01℃
temperature fluctuation ±0.1℃
heating rate 0.1~80℃/min
Constant temperature Room temperature -1000 ℃
incubation time 0-500min
Temperature control method PID temperature regulation
DSCresolution 0.01mW
DSCrange 0~±600mW
DSCsensitivity 0.1mW
Measurement range of balance 0.01mg-2g (optional range 10g, 50g, etc.)
Thermogravimetric resolution 0.01mg (optional resolution of 1ug, 0.1ug, etc.)
Gas control Nitrogen and oxygen gas control (instrument automatic switching), sealed furnace body, can achieve nitrogen protection, provide sealing test samples
working power supply AC220V/50Hz or customized
gas flow rate 0-200mL/min
gas pressure 0.2MPa
Optional crucible Aluminum crucible Φ 6.7 * 3mm, ceramic crucible Φ 6.7 * 3mm
Parameter standards Equipped with standard substances (indium, tin, zinc), users can calibrate temperature and enthalpy by themselves
Technical features:
1. Industrial grade 8-inch touch screen, displaying rich information.
2. The furnace body can be automatically lifted and lowered.
3. USB communication interface, strong universality, reliable and uninterrupted communication, supports self recovery connection function.
4. Water bath insulation system to isolate the influence of high temperature furnace body temperature on the weight of the balance.
5. Automatically switch between two atmospheric flow rates, with fast switching speed and short stabilization time. Simultaneously add a protective gas input.
6. Configuring a furnace body sealing system can achieve atmosphere protection environment and avoid air influence.
7. Improved installation process, all adopt mechanical fixation method, sample support rod is flexible and replaceable, and crucible can be selected with multiple models according to needs, making it convenient for users with different needs.
8. The flowmeter automatically switches between two gas flow rates, with fast switching speed and short stabilization time.
9. Weighing system using a dish scale, based on the principle of electromagnetic force balance.
10. The software is powerful and easy to operate.