1. Application
Usually used in conjunction with display instruments, recording instruments, electronic computers, etc., it outputs 4-20mA and directly measures the gas quality of liquids, vapors, and solid surface temperatures within the range of -200 ℃ to 1300 ℃ in the presence of explosives such as carbon and chlorine compounds in the production site.
2. Characteristics
Two line output 4-20mA, strong anti-interference ability;
Save the cost of compensating wires and installing temperature transmitters;
Safe and reliable, with a long service life;
Cold end temperature automatic compensation, non-linear correction circuit;
3. Working principle
Explosion proof thermocouples utilize the principle of gap explosion-proof. When an explosion occurs inside the chamber, it can be extinguished and cooled through the gap between the joint surfaces, preventing the full temperature of the flame after the explosion from being transmitted outside the chamber, thus achieving explosion-proof. The thermoelectric potential (resistance) generated by the thermocouple (thermistor) passes through the bridge of the temperature transmitter to generate an unbalanced signal. After amplification, it is converted into a 4-20mA DC electrical signal and sent to the working instrument, which displays the corresponding temperature value.
4. Product execution standards
IEC584
IEC751
IEC1515
JB/T5518-91
JB/T7391-94
5Integrated explosion-proof thermistor, integrated display meter, WZPBTemperature measurement range and allowable deviation
1) Temperature measurement range and tolerance of thermistor
Model number |
Dividing number |
Temperature measurement range ℃ |
accuracy class |
allowable deviation |
WZPB |
Pt100 |
-200-+500 |
Grade A, Grade B |
±( 0.15 + 0.002 ITI )
±( 0.30 + 0.005 ITI )
|
WZCB |
Cu50 Cu100 |
-50-+100 |
|
±( 0.30 + 0.005 ITI ) |
2) Thermocouple temperature measurement range and tolerance
Model number |
Dividing number |
Tolerance level |
I |
II |
Allowable deviation |
Temperature measurement range ℃ |
Allowable deviation |
Measuring range |
WRNB |
K |
±1.5℃ |
-40-375 |
±2.5℃ |
-40-333 |
± 0.004 ITI |
375-1000 |
0.0075% smile |
333-1200 |
WRMB |
N |
±1.5℃ |
-40-375 |
±2.5℃ |
-40-333 |
± 0.004 ITI |
375-1000 |
0.0075% smile |
333-1200 |
WREB |
E |
±1.5℃ |
-40-375 |
±1.5℃ |
-40-333 |
± 0.004 ITI |
375-800 |
0.004% smile |
333-900 |
WRFB |
J |
±1.5℃ |
-40-375 |
±1.5℃ |
-40-333 |
± 0.004 ITI |
375-750 |
0.004% smile |
333-750 |
WRCB |
T |
±1.5℃ |
-40-125 |
±1℃ |
-40-133 |
± 0.004 ITI |
125-350 |
0.0075% smile |
133-1000 |
6. Output signal: 4-20 mA , Load resistance 250, transmission wire resistance 100
7. Output method: Two line system
8. Accuracy level: Temperature transmitter accuracy level: 0.1 0.2 0.5 Display accuracy level: Analog indicator 2.5 level; Digital display level 0.5
9. Power supply:24V. DC±10 ℅
10. Protection level:IP65
11. Explosion proof grade:
Explosion proof type: d Ⅱ BT4 d Ⅱ CT5 d Ⅱ CT6 Intrinsic safety type: ia Ⅱ CT6
12. Insulation resistanceThe insulation resistance between the instrument output terminal and the housing should not be less than 50
13Integrated explosion-proof thermistor, integrated display meter, WZPBThermal response time:
When there is a step change in temperature, the time required for the current output signal of the instrument to change to 50% of the step change is usually represented by t0.5. When the step response stabilization time of the temperature transmitter does not exceed one-fifth of the thermocouple (resistance) thermal response stabilization time t0.5, the thermocouple (resistance) thermal response time is used as the thermal response time of the instrument; When the step thermal response stability time of the temperature transmitter does not exceed half of the thermocouple (resistance) thermal response stability time t0.5, the thermal response time of the temperature transmitter is used as the thermal response time of the instrument;
14. Basic errorThe basic error of the instrument should not exceed the combined error of the basic errors of the thermocouple (resistor) and temperature transmitter.
15. Working environment:
| Installation site level |
Temperature ℃ |
Relative humidity% |
Atmospheric pressure KPa |
| Cx1 |
-25-+55 |
5-95 |
86-106 |
| Cx2 |
-25-+70 |
| Cx3 |
-40-+80 |
16. Determination of support tube lengthThe working temperature of the temperature transmitter is the sum of the shell temperature rise caused by the support tube and the environment. The shell temperature rise caused by the support tube is shown in the following figure.
17. List of Evidence Collection:
| Explosion proof level |
Explosion proof certificate number |
Certification body |
| dIIBT4 |
GYB97151 |
NEPSI |
| dIIBT4 |
GYB97152 |
NEPSI |
| dⅡCT4 |
GYB97203 |
NEPSI |
| dⅡCT4 |
GYB97204 |
NEPSI |
| iaⅡCTT6 |
GYB99412 |
NEPSI |
| iaⅡCTT6 |
GYB99413 |
NEPSI |
18. Instrument wiring method
20Integrated explosion-proof thermistor, integrated display meter, WZPB modelSelection Table(For other sensor selection, please refer to the assembly of thermocouples/resistors or explosion-proof thermocouples/resistors)
| model |
graduation mark |
Temperature measurement range ℃ |
Thermal response time |
Explosion proof level |
Specifications |
| d |
L |
|
WRMB-140
WRMB-140G
|
K |
0-800 |
<90S |
dIIBT4
dⅡCT5
dⅡCT6
iaⅡCT6
|
Φ16 |
300X150
350X200
400X250
450X300
500X350
550X400
650X500
900X750
1150X1000
1650X1500
2150X2000
|
| <24S |
|
WRNB-140
WRNB-140G
|
N |
0-800 |
<90S |
| <24S |
|
WREB-140
WREB-140G
|
E |
0-600 |
<90S |
| <24S |
|
WRCB-140
WRCB-140G
|
T |
0-350 |
<90S |
| <24S |
|
WRFB-140
WRFB-140G
|
J |
0-55 |
<90S |
| <24S |
|
WZPB-141
WZPB-141G
|
Pt100 |
-200-500 |
<120S |
Φ12 |
| <24S |
|
WZCB-141
WZCB-141G
|
Cu50 |
-50-100 |
<120S |
| <24S |
1) The company can also customize various special models of thermocouples, thermistors, bimetallic thermometers, and flange sleeves according to samples or drawings.
2) Thermocouples of Class I and resistance of Class A shall be ordered according to the agreement;
3) The material of the protective tube is 1Cr18Ni9Ti, and other materials are ordered according to the agreement.
21. Selection Notice
Model A
B division number
Explosion proof grade C
Accuracy level of thermocouple (resistance)
Accuracy level of temperature transmitter (display)
F installation fixed form
Material of protective tube g
Length or insertion depth of h
|