- Phone
-
Address
No. 1090 Defu Road, Jiading District, Shanghai
Jiepu Instrument (Shanghai) Co., Ltd
No. 1090 Defu Road, Jiading District, Shanghai
InnoCon 6800OZ Online Ozone Analyzer Jepp Instrument
Measurement parameters: Ozone: 0.005-2.000ppm
Typical applications: drinking water/mineral water, disinfection industry, process water
characteristic
Two electrode system, Ampere membrane method principle
Wide power input, touch screen design
Large screen backlit LCD display for measurement values, temperature, and relay status
Chinese/English menu, easy to operate
Password protection to prevent unauthorized operations
New calibration steps tips can help reduce operational errors
2 x Programmable Hi/Lo Relay Output
Programmable automatic cleaning relay output
2 x isolated 4-20mA output
RS485 Modbus RTU communication
Data storage function, supports USB drive export
InnoCon6800OZ controller
Display: 4.3-inch color LCD touch screen display
Language: Chinese/English
Password: Setting mode: 0022, Calibration mode: 0011
Power supply: 110-240VAC, 50/60Hz;
Current output 1: isolated 4-20mA output, adjustable measurement value, maximum load 500 Ω
Current output 2: isolated 4-20mA output, adjustable measurement value, maximum load 500 Ω
Digital communication: RS485 Modbus RTU
Alarm output: 2 sets of programmable Hi/Lo contacts with hysteresis settings,5A/250VAC/30VDC
Cleaning the relay:Cleaning interval: 0.1-1000h, cleaning time: 1-1000s
Working temperature: 0~70.0 ℃
Storage temperature: -20~70.0 ℃
Protection level: IP65
Installation method: Wall mounted installation
Size: 200x180 × 90mm
Weight: ca.1Kg
InnoSens750 sensor
Measurement parameter: Ozone
Measurement range: 0.005-2.000ppm (mg/L)
Slope drift: approximately<-1%/month
Working temperature: Water sample temperature: 1-45 ℃, ambient temperature: 0-55 ℃
Temperature compensation: Built in temperature compensation
Work pressure: maximum 1 bar
Flow rate: approximately 15-30 liters per hour
PH operating range: 2-11pH
Polarization time: Approximately 1 hour for initial use
Response time: T90: approximately 15 seconds
Zero point adjustment: No need
Slope calibration: Single point calibration according to DPD method
Interference: CIO2 and Cl2
Material: PVC and AISI 316 Ti
Size: outer diameter 25mm, length 220mm
Storage: Sensor: 5-40 ℃, Electrolyte: 5-35 ℃, Membrane Cap: 5-40 ℃
Maintenance: The membrane cap replacement cycle is 1 year, and the electrolyte replacement cycle is 3-6 months (depending on the operating conditions)
InnoCon 6800OZ Online Ozone Analyzer Jepp Instrument