Overview

Infrared absorption spectrometer for detecting oil pollutants (such as petroleum hydrocarbons) 1. Detection target oil pollutants: including petroleum hydrocarbons (such as diesel, gasoline, lubricating oil), animal and vegetable fats, etc

Product Details

Infrared absorption spectrometer for detecting oil pollutants (such as petroleum hydrocarbons)

1. Target detection

  • oil and greaseIncluding petroleum hydrocarbons (such as diesel, gasoline, lubricants), animal and vegetable fats, etc.
  • key metricsThrough specific infrared absorption peaks (such asC-H bond stretching vibration absorption peak)Quantitative analysis of oil concentration.

2. Basic principles

  • Infrared absorption characteristics: In oil moleculesC-H bondIn the mid infrared band(3.4–3.5 μm, Namely 2940-2850 cm ⁻¹)Generate strong absorption peaks (stretching vibrations of methyl and methylene groups).
  • Quantitative basisAccording to Lambert Beer's law(Lambert-Beer Law)The absorbance is directly proportional to the concentration of oil:
    A=ε⋅c⋅l
    • A: absorbance
    • εMolar absorption coefficient
    • cOil concentration
    • l: Optical path length (sample cell thickness)

3. Testing process

  1. sample pretreatment:

    • extractionUse carbon tetrachloride (CCl ₄) to remove oil from water or solid samplesorTetrachloroethylene (C ₂ Cl ₄) * * extraction (both without absorption interference in the infrared region).
    • purify(Optional): Remove animal and vegetable oils (interferents) and retain petroleum hydrocarbons through a silicone column.
  2. instrument calibration:

    • Standard Curve MethodPrepare a series of concentrations using standard oils (such as n-hexadecane, isooctane, benzene, etc.), measure absorbance, and draw a standard curve.
    • Reference methodUsing the extraction solvent as a blank reference, eliminate solvent background interference.
  3. Sample measurement:

    • Inject the extraction solution into the sample pool and scan it3.4–3.5 μmBand absorbance.
    • Calculate the oil concentration based on the standard curve (unit: mg/L or mg/kg).

4. Key parameters and interference factors

  • detection limitUsually:0.1–1 mg/L(depending on extraction efficiency and instrument sensitivity).
  • Main interference:
    • Animal and vegetable oils and fatsOrganic compounds containing C-H bonds (to be separated by silica gel column).
    • suspended particulate matterFilter (<0.45 μ m filter membrane) to avoid scattering interference.
    • Solvent purityCarbon tetrachloride needs to be dealromatized to avoid residual benzene interference.

5. Instrument type

  • Non dispersive infrared (NDIR) spectrometer:
    • Specially designed for oil detection, with a fixed wavelength (3.4 μ m), simple structure, and low cost (commonly used for online monitoring).
  • Fourier Transform Infrared (FTIR) Spectrometer:
    • Full band scanning can distinguish oil components (such as petroleum hydrocarbons and vegetable oils), but the operation is complex and the cost is high (laboratory analysis).

6. Application scenarios

  • environmental monitoring:
    • Oil pollution detection in surface water, groundwater, and soil (such as emergency response to oil spills).
  • industrial wastewater:
    • Monitoring of oil content in wastewater from industries such as petrochemicals and mechanical processing.
  • Regulatory compliance:
    • comply withEPA 418.1(United States)ISO 9377-2(International) and other standards.

7. Comparison of advantages and disadvantages

advantage shortcoming
Strong specificity (C-H bond characteristic absorption) Pre processing is complex (requiring extraction and purification)
Fast (single detection<10 minutes) High solvent toxicity (strict protection is required for carbon tetrachloride)
Distinguishable between petroleum hydrocarbons and animal and vegetable oils (combined with purification) Affected by suspended solids and emulsified oil (requiring pre-treatment)

8. Comparison with other oil detection methods

method principle detection limit feature
Infrared absorption method C-H bond infrared absorption 0.1–1 mg/L High specificity, requiring solvent extraction
UV fluorescence method Aromatic UV fluorescence 0.01 mg/L High sensitivity, but only detects aromatic oils
gravimetric method Weigh after extraction 1–5 mg/L Easy to operate, but unable to distinguish oil components
Gas chromatography (GC) Separation and Quantification of Oil Components 0.01 mg/L High sensitivity, capable of component analysis, but time-consuming and costly

9. Precautions

  • Solvent selectionProhibit solvents containing C-H bonds (such as hexane) and prioritize the use of carbon tetrachloride (requires fume hood operation).
  • Sample preservationAfter extraction, the sample needs to be refrigerated in the dark (analyzed within 24 hours).
  • instrument maintenanceRegularly clean the sample pool