- Phone
-
Address
Room 901, No. 6, Lane 1011, Moyu South Road, Anting Town, Jiading District, Shanghai
Conner (Shanghai) Environmental Technology Co., Ltd
Room 901, No. 6, Lane 1011, Moyu South Road, Anting Town, Jiading District, Shanghai
Our company has launched an instrument that utilizes the principle of infrared spectroscopy absorption to measure low concentrations of carbon monoxide and carbon dioxide. Complies with the national standards of GB/T18204.2-2014 "Methods for hygiene inspection in public places - Part 2: Chemical pollutants", GBZ/T300.37-2017 "Determination of toxic substances in workplace air - Part 37: Carbon monoxide and carbon dioxide", and GB/T9801-1988 "Determination of carbon monoxide in air quality - Non dispersive infrared method".
This instrument complies with the national metrological verification regulations of JJG635-2011 "Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide", and its main technical indicators meet the technical requirements of national secondary instruments. It can obtain a verification certificate from the Chinese Academy of Metrology (verification conclusion: qualified, this instrument meets the secondary technical requirements).
At the bottomDetection principle: Non dispersive infrared gas analysis method/Non dispersive infrared method (national standard)
At the bottomDetecting gases: Carbon monoxide (CO), carbon dioxide (CO2) in the air
At the bottomDetection method: Built in pump suction
At the bottomMeasurement range: 0~50.0 × 10-6CO,0~0.500%
At the bottomConcentration display ppm, mg/m3 automatic conversion temperature and humidity display
At the bottomResolution: 0.1 × 10-6 CO;0.001% CO2
At the bottomColor touch screen operation, with 485 data interface to automatically store 5000 sets of measurement data
At the bottomAutomatic zero point correction technology, convenient for users to perform zero point correction in different seasons and times
At the bottomLinearity: ≤± 2% full-scale reproducibility: ≤ 1% full-scale
At the bottomZero drift: ≤± 2% full-scale/h span drift: ≤± 2% full-scale/3h
At the bottomInterference error: 1250mg/m3CO ≤± 0.3% full range
At the bottomResponse time: CO: t0~t90≤45S; CO2:t0~t90≤15S、 Preheating time: 30 minutes
At the bottomFlow range: (0.5-2.0) L/min
At the bottomPower supply: AC/DC dual-use, 220AVC (± 10%) or built-in rechargeable battery