Features:
1. Digital sensor, directly outputting Rs-485 digital signal
2. 90 ° angle scattered light principle, built-in temperature can automatically compensate
3. Fiber optic structure with strong resistance to external light interference
4. Infrared LED light source, with added filter design, resistant to light source interference, and good stability
5. Surface anti-corrosion and passivation treatment
6. Low power consumption of sensors and anti-interference design of internal circuits
Nameparameter
Output signalSupport Rs-485 output
Measurement method90 ° scattered light method
Range0-200NTU or 0-1000NTU or 0-3000NTU
Accuracy± 5% indication or ± 3NTU, whichever is greater (0-200NTU)± 5% indication or ± 5NTU, whichever is greater (0-1000NTU)± 5% indication or ± 10NTU, whichever is greater (0-3000NTU)±0.5℃
Resolution0.1NTU
Working conditions0~50℃
Pressure range<0.2MPa
Calibration methodTwo-point calibration
Response time30 seconds (T90)
Temperature compensationAutomatic temperature compensation (Pt1000)
Power supply12 or 24V
Protection levelIP68

Maintenance and precautions
1. Maintenance
It is recommended to clean the attachment of the sensor measurement probe every 30 days; Avoid using hard objects to scratch the light guide part of the measuring probe during cleaning; Please wipe with a soft damp cloth.
It is recommended to clean the outer surface of the sensor with water flow. If there are still debris residues, please wipe them with a damp soft cloth.
2. Precautions
Installation measurement: Avoid conducting installation measurement in areas with turbulent water flow to reduce the impact of water bubbles on the measurement. Keep the measuring probe 5cm away from the bottom.