- Phone
-
Address
Room 1204, Regis Ginza Enterprise Center, 1388 Shaanxi North Road, Shanghai
Grant Instruments (Cambridge) Limited, UK
Room 1204, Regis Ginza Enterprise Center, 1388 Shaanxi North Road, Shanghai
Imported low-temperature circulating water bathIt is a high-performance, programmable temperature controller with a full-color QVGA TFT display for actual and set temperatures. The intuitive screen icons and menus allow for quick setting of multiple parameters.
TX150 is programmable and can be remotely controlled by a PC through a USB interface for data storage. Through Labwise? The software can set all parameters, including programming, and display the operating conditions of the controller in real time, while recording data for future retrieval and analysis.
Imported low-temperature circulating water bathcharacteristic
Powerful built-in circulation pump: maximum flow rate of 18L/minute (adjustable), maximum pressure of 310mbar
Color screen QVGA TFT display
Add 3 adjustable preset temperatures to the operating set point
Reaction timing, 99 hours, 59 minutes audible alarm and heater control, timer function
The temperature range and heating power are safely adjusted according to the type of liquid selected
fault display
Programming Relay
User recalibration of internal and external probes at two points (TXPEP. TXSEP)
The temperature compensation function is done * temperature
USB port
Optional LabwiseTMComputer program settings, data recording, and real-time graphics software
Can edit one 30 step program
Language options: English, French, German, Spanish, and Italian
Low liquid level protection function, achieved through float switch
Adjustable over temperature cut-off protection sound alarm
Equipped with convenient clamps for easy replacement of the bath
5 types of cooling baths (R series) are available for different conditions, working capacities, and cooling power.
Cooling/Heating Range: -47 to 100 ℃ (depending on the selected cooling unit)
Stability: ± 0.1 ° C
Note: All refrigeration units can be turned off independently of the thermostat controller. The maximum operating temperature for all refrigeration units is 100 ℃
technical parameters
| Grant Optima ™ Technical parameters of constant temperature controller | ||||||
| ●=Standard configuration | Universal digital type | High performance digital type | ||||
| T100 | TC120 | TX150 | TXF200 | |||
| Stability (DIN12876) | Water at 10 ℃ | ℃ | ±0.1 | ±0.1 | ±0.1 | ±0.1 |
| 50% water, 50% ethylene glycol | @10℃ | ℃ | - | ±0.1 | ±0.1 | ±0.1 |
| Uniformity (DIN12876) | Water at 10 ℃ | ℃ | ±0.1 | ±0.1 | ±0.1 | ±0.1 |
| 50% water, 50% ethylene glycol | @10℃ | ℃ | - | ±0.1 | ±0.1 | |
| set precision | ℃ | 0.1 | 0.1 | 0.1 (0.01 using Labwise) ™ Software) | ||
| display mode | 4-digit LED | Color screen QVGA TFT | ||||
| timer function | - | 1 to 9999 mins | 1min to 99h59mins | |||
| Number of preset temperatures | 3 | 3 | 3 | 3 | ||
| Two-point calibration | ● | ● | ● | ● | ||
| Temperature compensation function | - | - | ● | ● | ||
| External temperature probes (TXPEP, TXSEP) | - | - | ● | ● | ||
| communication interface | - | - | USB&RS232 | USB&RS232 | ||
| Programming mode | - | - | PC remote control/1 programming 30 steps | Interface control/remote PC control/10 programming 100 steps | ||
| relay | - | - | 1 | 1 | ||
| safety | over-temperature protection | fixed | Adjustable disconnect | |||
| Low liquid level protection - floating on light | ● | ● | ● | ● | ||
| Alarm (can set a relay control switch) | - | - | high and low temperature | high and low temperature | ||
| heating power | 230V KW | 1.3 | 1.3 | 1.9 | 1.9 | |
| 120V KW | 1.4 | 1.4 | 1.4 | 1.4 | ||
| total power | 230V KW | 1.4(50-60Hz) | 1.4(50Hz) | 2.0(50Hz) | 2.0(50-60Hz) | |
| 120V KW | 1.5(50-60Hz) | 1.5(50-60Hz) | 1.5(50-60Hz) | 1.5(50-60Hz) | ||
| Above the height of the edge of the bath | mm | 200 | 200 | 200 | 200 | |
| Depth below the edge of the bath | mm | 135 | 135 | 135 | 135 | |
| Grant Optima ™ Thermostatic circulation pump (built-in) | ||||||
| The biggest pressure | Water bar | - | 210 | 310 | 530 | |
| Zui high flow velocity | Water L/min | - | 16 | 18 | 23 (adjustable flow rate) | |
| Pipeline diameter | In/out mm | - | 6/11 | 6/11 | 6/11 | |
| R Series Refrigerator - Model and Technical Parameters | |||||||
|---|---|---|---|---|---|---|---|
| Grant Optima ™ circulator | |||||||
| ●=Standard configuration | R1 | R2 | R3 | R4 | R5 | ||
| Relay control function (refrigeration unit) | - | ● | ● | ● | ● | ||
| refrigerant | R134a | R134a | R134a | R134a | R134a | ||
| drain outlet | - | ● | ● | ● | ● | ||
| Over temperature protection function | 100 ℃ restriction | ● | ● | ● | ● | ● | |
| Waterproof icing function | ● | ● | ● | ● | ● | ||
| High pressure refrigeration switch | 27bar | - | - | - | ● | ● | |
| Cooling power, room temperature 20 ℃ | @20℃ | W | 250 | 250 | 160 | 900 | 1100 |
| @0℃ | W | 140 | 140 | 150 | 500 | 1050 | |
| @-10℃ | W | 100 | 100 | 120 | 300 | 800 | |
| @-20℃ | W | 35 | 35 | 50 | 180 | 580 | |
| @-30℃ | W | - | - | 5 | 40 | 370 | |
| @-40℃ | W | - | - | - | - | 130 | |
| @-47℃ | W | - | - | - | - | 25 | |
| power | 230V | W | 334(50Hz)* | 334(50Hz)* | 354(50Hz)* | 850(50Hz) | 1400(50Hz) |
| 120V | W | 328(50-60Hz) | 328(50-60Hz) | 370(60Hz) | 780(60Hz) | - | |
| Electromagnetic Radiation Registration | class | B | B | B | B | B | |
*Optima ™ The temperature controller and accessory pump can be powered through the R1, RW, and R3 220-240V refrigerator back sockets. The power of the refrigerator is increased by an additional 2kW
**The optional VTP pump will bring an additional two liang to the bath, so without refrigeration accessories, the lowest temperature that the bath can reach will increase.
Attention: When ordering a VTP pump, please indicate the model of the Grant bath that comes with it.