The IRO-300 PLUS fully automatic infrared spectrophotometer consists of a fully automatic operating software system, infrared spectroscopy system, and pre-processing extraction system. Fully automated, without the need for operators to come into contact with tetrachloroethylene, including automatic sampling, extraction, water and impurity removal, measurement, cleaning, drainage, and data storage.
The IRO-300plus fully automatic infrared oil analyzer consists of a fully automatic operating software system, an infrared spectroscopy system, and a pre-processing extraction system. Fully automated, without the need for operators to come into contact with tetrachloroethylene, including automatic sampling, extraction, water and impurity removal, measurement, cleaning, drainage, and data storage.
IRO-300 PLUS Fully Automatic Infrared Spectrometer for Oil AnalysisTechnical Features
1. Fully automatic one click completion of sample injection, extraction, water filtration, measurement, drainage, and cleaning, capable of extracting 1-12 water samples continuously at once.
2. High precision automatic injector: reagent 0-50ml can be set arbitrarily, automatic dilution for excessive concentration, automatic preparation of quality control samples, and automatic one click calibration for curve making.
3. Volume measurement: Non contact instruments automatically measure the volume of water samples, with a volume measurement error of less than 2%.
4. The use of infrared oil standard filters allows for one click calibration, multiple calibrations without the need for reagents, and automatic detection and calibration at both single and multiple points.
5. Dual membrane oil-water separation technology can process emulsified water samples, high chromaticity, and turbidity water samples. One membrane can be used up to 100 times, and there is an audible and visual alarm when the membrane breaks.
6. Built in 12 inch touch computer, Windows full system control, can also be remotely operated and monitored through a mobile app, and data can be uploaded to the laboratory data platform.
7. The pre-treatment of the instrument adopts anti-corrosion material semi-circular push-pull technology for door opening and closing, as well as a 360 degree automatic rotating turntable. The water sample is easy to place and has good sealing performance.
8. One click switch between semi-automatic and fully automatic operation.
9. The instrument automatically displays the remaining amount of magnesium silicate and can automatically replace magnesium silicate infinitely without replacing the magnesium silicate column, which complies with the national standard.
10. Scanning of three wavenumber characteristic peak spectra, automatic scaling of spectra, and traceability verification of sample concentration through absorption peak scanning spectra.
11. The manufacturer is equipped with a dedicated sampling bottle with a concave bottom for waste discharge and no residual reagents. The chain type sample turntable is compatible with commercially available 500ml and 1000ml wide mouth bottles.
12. Fault self check: Automatically detect pre-processing and host faults before startup, and sound and light alarms for faults.
13. Automatic separation and collection of wastewater and waste tetrachloroethylene, automatic detection and display of reagent residue, and upgrading of automatic recycling and reuse of waste liquid.
IRO-300 PLUS Fully Automatic Infrared Spectrometer for Oil Analysistechnical indicators
Project Description
The detection limit of the instrument is DL<0.02mg/L (measure 11 times blank and calculate 3 times standard deviation)
Wave number range 3400cm-1 to 2400cm-1 (i.e. 2941nm to 4167nm)
The absorbance range is 0.0000-2.0000 AU (i.e. transmittance of 100-1% T)
Detected concentration 0.001 (oil concentration in water)
The highest measured concentration ranges from 0.001 to 100000mg/L
Basic measurement range: 0.001-200mg/L
The relative standard deviation (RSD) of repeatability is ≤ 0.5% (11 measurements of 20-100mg/L oil standard sample)
Measurement accuracy error<± 2%
Baseline stability: baseline drift less than 0.005AU/60min
Scanning speed of 30 seconds per scan, fixed point multiple scans for 10 seconds, non dispersive infrared method for 2 seconds,
Correlation coefficient R>0.9995
Water sample volume 5ml-600ml
Grade A resolution<3%
Test sample size: Continuously test 1-12 samples
Automatic detection time for a single sample is 2-8 minutes
Wave number accuracy and wave number repeatability ± 1cm-1
Net weight of the host is 22kg
The overall dimensions of the machine are 100cm × 55cm × 65cm (length, width, height)