19431Human walking static electricity tester
19431Human walking static electricity testerProduct Features
Test the static voltage generated by human walking to evaluate anti-static shoes-Comprehensive static dissipation performance of flooring
Reading mode (without saving data), curve mode (saving data)
Range:0~+/-1050V
Resolution:1V
Test accuracy:+/-5V (within 100V),+/-5% (100V-1050V)
Input impedance:1014Ω
Sampling rate: Reading mode-5Hz (0.2 second readings), curve mode -200Hz (0.005 second readings)
Testing time: longest2 minutes and 30 seconds
The instrument can be saved10 curve graphs that can be output to a computer
19431 Human walking static electricity teststandbyAutomatic shutdown after 10 minutes of inactivity
19431 Human walking static electricity testWorking power supply:6 AA batteries
Order Number
19431 Human walking static electricity testKit, including the following items
test table
Grip electrode and connecting wire
Data cable(8针端子-USB端子)
Grounding wire(5 meters long)
Grounding plug (item number)09838C)
Installation, operation, and maintenance instructions for a body voltmeter. The DeSCO 19431 body voltmeter is a portable measurement system that can analyze the accumulated charge level of the human body according to the ANSI/ESD STM 97.2 standard. The measurement of the body voltmeter can be recorded using internal memory. The recorded measurement results can be exported and further viewed on a PC. The measurement curve can be displayed on the LCD screen for quick evaluation. Every human voltmeter is calibrated andThe accreditation procedures and standards can be traced back to the National Institute of Standards and Technology (NIST). In October 2023, a compliance verification plan should be developed to ensure that the organization meets the requirements of the plan. Process monitoring (measurement) should be conducted based on the compliance verification plan, which should determine the technical requirements to be verified, measurement limits, and frequency of verification. The compliance verification plan should record the test methods used for process monitoring and measurement. If the organization uses different testing methods to replace the testing methods of this standard, it should be able to demonstrate that the results obtained are relevant to the referenced standard. When designing testing methods for projects not covered by this standard, these methods should be fully documented, including corresponding testing limits. Compliance verification records should be established and maintained to provide evidence of compliance with technical requirements. The selected testing equipment should Capable of conducting measurements as specified in the compliance verification plan. ”(EN 61340-5-1 clause 5.2.4) Package 1 body voltmeter 1 walking test probe 1 black 8-pin DIN to USB cable, 1.8 m long 1 black grounding wire with crocodile clip, 5 m long 1 hanging rope with hook