HP9920 High speed Insulation Resistance Tester: $n is an instrument used to test the short circuit and insulation resistance of battery cells. Compliant with IEC 61010 safety standard. The main application of lithium batteries is to apply high voltage testing to the battery cells before injecting electrolyte, to determine insulation faults such as open circuit, short circuit, and micro short circuit of the battery cells. And it has abnormal detection function and short circuit testing function, which can quickly determine and test for up to 50ms.
HP9920 High speed Insulation Resistance Tester-Performance characteristics:
Capacitive load enables fast charging and discharging; Stable testing without the need for additional shielded wires; The test piece connection detection function prevents open circuit misjudgment; Short circuit detection before testing to prevent high-voltage breakdown; Continuous testing mode, PASS termination mode; Termination mode, forced termination mode; Rich interfaces, suitable for pipeline sorting.
| HP9920 Lithium Battery Cell Short Circuit and Insulation Resistance Tester: It is an instrument used to test the short circuit and insulation resistance of battery cells. Compliant with IEC 61010 safety standard. The main application of lithium batteries is to apply high-voltage pulse testing to the battery cells before injecting electrolyte, to determine insulation faults such as open circuit, short circuit, and micro short circuit of the battery cells. And it has abnormal detection function and short circuit testing function, which can quickly determine and test for up to 50ms. |
| HP9920 High speed Insulation Resistance Tester-Main features: |
- Free setting of test voltage: 10-1000 V, step by 1V, select the appropriate test voltage according to the requirements of testing different specifications of battery cells;
- Quick judgment: testing time up to 50ms, quick comparison function, upper and lower limit setting judgment;
- Contact and short circuit anomaly detection: using four terminal contact detection method, with contact/short circuit anomaly detection function, to prevent misjudgment caused by poor contact and to prevent defects caused by micro short circuits;
- Rapid discharge: Automatically release residual voltage at high speed.
|
| HP9920 Insulation Resistance Tester - Technical Specifications: |
| Function | Insulation resistance tester | | Display | 3.5-inch high-resolution TFT screen display | | test voltage | Voltage regulation range of 10-1000 V, step by 1V (resolution) | | range | Large display | Accuracy guarantee range | Basic accuracy | | 25 V ≦ V < 100 V | 2 MΩ | 4.000 MΩ | 0.000 MΩ~2.000 MΩ | ±2% rdg. ±5 dgt. | | 20 MΩ | 40.00 MΩ | 1.90 MΩ~20.00 MΩ | ±2% rdg. ±5 dgt. | | 200 MΩ | 400.0 MΩ | 19.0 MΩ~200.0 MΩ | ±5% rdg. ±5 dgt. | | 100 V ≦ V < 500 V | 2 MΩ | 4.000 MΩ | 0.000 MΩ~2.000 MΩ | ±2% rdg. ±5 dgt. | | 20 MΩ | 40.00 MΩ | 1.90 MΩ~20.00 MΩ | ±2% rdg. ±5 dgt. | | 200 MΩ | 400.0 MΩ | 19.0 MΩ~200.0 MΩ | ±5% rdg. ±5 dgt. | | 2000 MΩ | 4000 MΩ | 190 MΩ~2000 MΩ | ±5% rdg. ±5 dgt. | | 500 V ≦ V ≦ 1000 V | 2 MΩ | 4.000 MΩ | 0.000 MΩ~2.000 MΩ | ±2% rdg. ±5 dgt. | | 20 MΩ | 40.00 MΩ | 1.90 MΩ~20.00 MΩ | ±2% rdg. ±5 dgt. | | 200 MΩ | 400.0 MΩ | 19.0 MΩ~200.0 MΩ | ±2% rdg. ±5 dgt. | | 4000 MΩ | 9990 MΩ | 4010 MΩ~9990 MΩ | ±25% rdg. | | range switching | manual | | High output current | 1.8mA | | test speed | Fast/Slow; The small testing cycle only takes 50ms | | test mode | Measurement value ≥ upper limit value; Upper limit value>measured value>lower limit value; Measurement value ≤ lower limit value | | judgment processing | Beep sound, indicator light: qualified (IN), unqualified (upper exceeding HI, lower exceeding LO), EXT. I/O output | | timer | 0.100 s ~ 999.999 s | | Abnormal connection display | Open circuit "ContHi" at the high voltage end, open circuit "ContLo" at the low voltage end, and open circuit "ContHL" at both output ends | | Short circuit abnormal display | “SHORT” | | Range overrun display | Exceeding the range by "Underer. F" and "OVER. F" respectively | | discharge current | Quickly release residual voltage of the tested product | | Equipped with interfaces | Optional: (RS232C, LAN Ethernet interface communication; Handler (I/O port); EX.SW external switch interface, used to trigger or terminate testing; Analog output interface. )Control or measure values, and obtain judgment results of this instrument through a computer or PLC. It also has the function of automatically sending measurement values or judgment results after each test is completed. | | Operating temperature/humidity | 0 ℃ to 40 ℃; Below 80% RH (no condensation) | | Storage temperature/humidity | -10 ℃ to 60 ℃; Below 80% RH (no condensation) | | operating environment | Indoor, altitude 2000 meters | | power supply | Voltage: 100V~240V AC Frequency: 50Hz/60Hz | | Power consumption | 25VA | | Size | 325mm x 215mm x 96 mm | | Weight | 2Kg | |