- Phone
-
Address
New City Science and Technology Industrial Park, Xincheng District, Xi'an City
Shaanxi Fusaisi Instrument Co., Ltd
New City Science and Technology Industrial Park, Xincheng District, Xi'an City
![]()

Handheld smoke detector for environmental protection and haze detectionInstrument principle:Comprehensive Optical Analysis of Optical Path
characteristic:
„ Comprehensive Optics Professional Optical Suspended Matter Testing Technology
„ Equipped with intelligent instrumentsCan meet various testing requirements
„ Featuring small size, light weight, and easy portability
„ Fast and accurate measurement, digital display, high sensitivity, and stable performance
Application of smoke detector:
① Environmental protection smoke detection② Analysis of Foggy in Dusty Spaces③ Tunnel smoke concentration detection④ Gas purification and oil mist detection⑤ Atomization concentration detection
⑥ Mist detection
Portable smoke concentration detectorTunnel smoke concentration meter,Portable smoke detection instrumentSmoke experimental instrument
Smoke tester function:
Portable smoke concentration detectorTunnel smoke concentration meter,Portable smoke detection instrumentSmoke experimental instrument
‹ During normal testingSimply insert the sensor into the sample and turn on the power switch The test value will be displayed on the LCD monitor within a few seconds ‹ Instantaneous test values can be used<HOLD>key to lock display
‹ Can store test valuesEach calibration table corresponds to 24 data recording functions‹ Readable, average, and delete records, can configure the printer to print test records‹ The instrument has a backlightIn dark environments, lights can be turned on for illumination
Handheld smoke detector for environmental protection and haze detectionInstrument technical specifications:
Measurement principle:Measurement of relative mass concentration of transparent and opaque suspended solids through flux testing
Measurement object:Detection of various haze levels in atmospheric environment
Measurement range:0.8-800mg/M3(Referring to water mist, the higher the density, the larger the range. Particle size is different.)Cloth can affect numerical values. Dark absorbing substances generally reduce the range)
Measurement time:1s. Professional regulations require taking the averageThe function of recording data and then averaging can be used to meet the requirements
Measurement accuracy:<2% or0.8 g/M3(Water mist) ,2KS two-point calibration;
monitor:16x2 LCD liquid crystal display
Serial communication:RS232/485
Ambient temperature-20 to 60℃
power supplyLithium iron battery type: can be used continuously12 hours, equipped with a 220-12V charger
Size:Handheld watch:98X150X30mm; Sensor: D50x360 mm
Total weight:0.7Kg