E-mail
sales@meas-elec.com
-
Phone
18328702053
-
Address
Room 1, 4th Floor, Unit 1, Building 4, No. 166 Wuxing Fourth Road, Wuhou District, Chengdu City
Chengdu Hongruitao Technology Co., Ltd
sales@meas-elec.com
18328702053
Room 1, 4th Floor, Unit 1, Building 4, No. 166 Wuxing Fourth Road, Wuhou District, Chengdu City
Chengdu Hongrui TaoHT-FX100-H2 Oxygen Hydrogen Analyzer
HT-FX100Oxygen hydrogen transmitterUsed for measuring hydrogen content in ammonia decomposition and protective atmosphere equipment in environments with explosion-proof grade requirements of Exd IIC T6, and for online measurement of hydrogen content in welding and cutting proportioning gases; Online monitoring of hydrogen content in hydrogenation purification nitrogen generator.

Instrument features:
a.Segment LCDDisplay, intuitive display, simple and convenient operation;
b.adopt
thermal conductivitysensorOr fuel cellssensor,haveHigh testing accuracy, long calibration interval, and long service lifeOther characteristics;
c.Menu lock function to prevent accidental changesinstrumentParameters that affect its performance;
d.In explosion-proof situations, it can be achieved throughisolatedMagnetic pen completes menu settings for easy on-site operation;
e.When the power is turned off, the service life of the sensor is not lost (thermal conductivity principle);
f.Automatic temperature compensation function,Eliminating temperature fluctuationsmeasurementvalueThe impact;
g.The instrument supports multi-point segmented linearitycalibration,satisfyFull rangescopeconcentrationaccurate measurement;
h.The test chamber airway is specially designed with embedded polymer membrane, which can effectively prevent the influence of liquids and interfering media on the sensor;
Thermal conductivity principle/fuel cellTesting principle (principle optional).
Chengdu Hongrui TaoHT-FX100-H2 Oxygen Hydrogen Analyzer
Technical Specifications:
lMeasurement principle: Thermal conductivity principle/Fuel cell (principle optional)
ldisplay mode:segment codedisplay
lMeasurement range: Thermal conductivity principle:0~30.00%/50.00%/99.99% H2(Range optional)
fuel cell:0~10PPm/100PPm/2000PPm H2(Range optional)
ldivide distinguish Rate: Thermal conductivity principle:0.01%
fuel cell:1PPm
lMeasurement accuracy: thermal conductivityPrinciple:≤±1.0%FS
fuel cell:≤±1.5%FS
lheavy again sex:≤±1.0%FS
lresponse time: Thermal conductivityPrinciple:T90≤30S
fuel cell:T90≤60S
loutputinterface:4-20mA. DC(Non isolated output,load resistor<1KΩ)
1Programmable dry contact type passive alarm output,Contact CapacityAC220V/2A