HS-D-5800P Butler Service $r $nXccelerator Accelerator Service $r $n Customizable Exclusive Software Version $r $n Quick Response After sales Channel
HS-D-5800P Butler style Service
Aijer Fenomi has always been committed to providing customers with high value-added after-sales and technical support services. Adhering to the customer-centric concept, we always insist on providing after-sales service with a quick response within 2 hours, eliminating any worries you may have when using it. In addition, the supply of commonly used spare parts is guaranteed by a professional supply chain team, which can quickly solve common instrument usage and maintenance problems.
The HS-D-5800P preparation software is provided with continuous technical support and upgrade services by professional software engineers from the Aijer Fenomi team, and will continue to be committed to continuously meeting the needs of practical users in preparation and purification. Moreover, our local R&D team can provide personalized software customization, upgrade, and iteration services for specific customers based on their special usage habits and requirements, providing you with exclusive software versions and serving each customer more thoughtfully.
In conjunction with the Aijer Fenomi Xccelerator acceleration service, more than 30 experienced application scientists from PhenoLogix in Shanghai, Tianjin, and the United States will utilize their years of laboratory experience to provide you with technical services for quickly establishing purification methods, helping you choose suitable consumables to match with HS-D-5800P to achieve high efficiencygoodThe preparation and purification performance of the CMC accelerated package form a complete scheme.
Introduction to the functions and parameters of each module
collector
▲ Independent collection bed mode, non-interference, compact design, supports up to 5 test tube rack positions;
▲ Support 13 mm, 16 mm, and 18 mm test tube racks, freely choose according to usage habits;
Accurate pipeline design to achieve smaller delay volume, reduce collection dead volume, and minimize collection inaccuracies caused by sample diffusion in the flow cell;
▲ Adopting high-precision three-way valve with high pressure resistance to switch and collect,especiallyThe collection switching design ensures accurate and leak free switching of collection positions while maintaining the independence of the waste liquid channel, ensuring the recovery rate.
| collector |
AX-C501 High Pressure Preparation Series Fractional Collector |
| Dimensions and Weight |
680×837×658 mm (H×W×D),45 Kg |
| power supply |
100-240V AC, Power 75 W |
| capacity |
13 × 100 mm, 9 mL, 96 wells, 5 test tube racks 16 × 100 mm, 15 mL, 75 wells, 5 test tube racks 18 × 150 mm, 25 mL, 70 wells, 5 test tube racks (The maximum height of the test tube supports 180 mm) |
| switching speed |
The fastest speed is 0.1 seconds |
| flow range |
0-150 mL/min |
Introduction to the functions and parameters of each module
detector
▲ Dual path optical system, achieving dual wavelength detection;
▲ The monochromator adopts imported high-precision control, which improves wavelength accuracy;
▲ Independent research and development of optical path and design of quick detachable circulation pool.
| name |
HP-Q-UV2080 Preparation Type UV Variable Wavelength Detector |
| Dimensions and Weight |
172×240×450 mm (H×W×D),18 Kg |
| power supply |
100-240V AC, Power 75 W |
| wavelength range |
200-800 nm, Increment of 1 nm, deuterium and tungsten lamp light sources |
| wavelength accuracy |
1 nm |
| Wavelength precision |
± 0.1 nm |
| noise |
±0.5×10-5 AU |
| drift |
2×10-4 AU |
| Flow pool volume |
10 μL |
| Flow pool path length |
2.1 mm |
| Testing Scope |
0-2 Au |
pump
▲ Dual plunger parallel connection, floating plunger design, ensuring accuracy while effectively extending the service life of the sealing ring;
▲ Reserved plunger cleaning channel, which can effectively clean the high salt buffer residue on the back of the sealing ring without disassembling the pump head. The embedded pump head design is easy to disassemble and reset, making maintenance simpler;
▲ The motor uses a servo motor with a higher speed and is processed by reducing the speed and increasing the torque through two-stage transmission, further improving the stability of the pump gradient and equal flow rate;
Traditional slider linear bearings experience increased axial force on the slider and tangential force generated by the pressure angle of the cam when the pressure increases, making it difficult for engineering plastic linear bearings to ensure the stability of the slider. This product uses ball type linear bearings to solve stability issues and further extend the service life of the sealing ring;
Traditional pump heads are machined to ensure concentricity requirements. Compared to floating high-pressure sealing ring seat sleeves, the sealing and concentricity effects are better, and the stability of the pump is stronger.
| name |
HP-Q-P150 150 mL Preparation High Pressure Infusion Pump |
| Infusion system |
Dual plunger parallel mode |
| Dimensions and Weight |
172 × 278 × 450 mm (H × W × D), 18.5 Kg (including mixer) |
| power supply |
100-240V AC, Power 300 W (single pump) |
| Pipeline connection |
1/16 "standard pipeline connection |
| pressure pulsation |
≤0.5 Mpa |
| pressure range |
≤ 40 Mpa (316 L pump head) |
| flow range |
0.01~150 mL/min, Incremental 0.01 mL/min |
| Flow accuracy |
±1% |
| Gradient error |
≤1.5% |
| Traffic repeatability |
RSD≤1% |
Integrated power box
| Integrated power box |
ADC-PCB-5800P High Voltage Liquid Phase System Power Supply and Communication Integration Module |
| Dimensions and Weight |
172×240×461 mm (H×W×D), 10 Kg |
| power supply |
110-220V AC, Power 100 W, including column switching module 200 W |
| IoT module |
Support SIM cards from China Mobile, China Unicom, and China Telecom |
Column switching (column bracket)
| Column switching |
ACQ-501 automatic preparation column switching module |
| Dimensions and Weight |
Column bracket: 460 × 520 × 190 mm (H × W × D), 5 Kg Switching Module (including Integrated Power Box): 172×240×461 mm (H×W×D), 13 Kg |
| power supply |
110-220V AC, Power 100 W, including integrated power box 200 W |
| number of channels |
5 通道 |
sampler
▲ Three types of test tube rack designs, supporting 5 mL, 9 mL, and 15 mL sample tubes with graduations for more flexible matching as needed;
▲ Independent automatic injection module, which does not interfere with the collection time, can inject separately or achieve repeated superimposed injection;
The injection volume can be adjusted from 100 μ L to 10 mL (± 1 μ L), and the injection speed can be controlled within 30 seconds from small volume to amplification, achieving rapid injection;
▲ Sample residual is less than 0.01%, and higher reproducibility can also be achieved for complex samples;
▲ In response to the customer's demand for larger injection volumes, a switch port beyond the syringe range has been reserved for expansion, from 100 μ L-5 mL injection volume to 100 μ L-20 mL injection volume.
| sampler |
ATS-X01-H10 High Voltage Automatic Sample Injector |
| Dimensions and Weight |
400×340×525 mm (H×W×D),35 Kg |
| power supply |
110-240V AC, Power 75 W |
| injection volume |
100 μ L-5 mL (5 mL quantitative loop, Minimum increment of 1 μ L, expandable to 20 mL) |
| Repeatability of injection flow rate |
RSD≤1% |
| Accuracy of injection flow rate |
±2% |
| Injection cycle time |
5 mL standard speed injection time is 32 seconds 10 mL standard speed injection time is 50 seconds |
| Injection dead volume |
310 μL |
| Residual effect |
≤0.01% |
| Sample rack type |
10 × 75 mm, 5 mL, 112 wells 13 × 100 mm, 9 mL, 72 wells 16 × 100 mm, 15 mL, 50 wells 30 × 100 mm, 40 mL, 6 wells |
| Applicable pressure |
≤40 Mpa |
| cleaning pump |
Support cleaning of the inner and outer walls of the injection needle |
| Sports mechanism mode |
X-Y-Z |