Product Introduction: The fully automatic fluorine and chlorine analyzer features PID control of furnace flow, low-temperature drift components, high temperature control accuracy, and extended use time of silicon carbon tubes

Product Details

Product Introduction

The functional characteristics of the fully automatic fluorine and chlorine analyzer are:
Adopting PID control furnace flow, low-temperature drift components, high temperature control accuracy, and prolonging the service life of silicon carbon tubes.
The evaporation rate is controlled by solid-state relays, and the voltage is continuously adjustable from 0-220V.
Configure a digital ion meter, a precision automatic titrator, and a computer program to automatically collect various data and curves. The measurement results are automatically calculated.
It can automatically titrate and set the titration speed and volume by oneself.
Can automatically send samples.
At the same time, our company can also develop various titration programs for users with different usage requirements to achieve computerized titration operations and improve assay accuracy.

Parameters of fully automatic fluorine chlorine instrument:
Temperature control accuracy: 1100 ± 2 ℃
Heating time: Heat up to 1100 ± 2 ℃ within 30 minutes
Power supply: AC220V ± 10%, 50HZ
High temperature hydrolysis time: 30 minutes with fluorine; Chlorine for 30 minutes;
Titration time (excluding hydrolysis time):
Fluorine: 3 minutes (excluding sample processing time, sample processing time is 30 minutes)
Chlorine: 3 minutes (including sample processing time)
Measurement accuracy: Fluorine: ≤ 150ug/g ± 15; >150ug/g±10%;
Chlorine: ± 0.01
Measurement range: Fluorine: 10ug/g-2000ug/g
Chlorine: 10ug/g-2000ug/g
Test method: High temperature combustion hydrolysis method.
Whole machine power: ≤ 3.5KW.
The stirring speed is continuously adjustable and displayed digitally. The computer automatically collects measurement data and calculates automatically.