OtalPerm is a better choice for gas permeability testing, making experiments easier while reducing costs and saving space. You can use software to plan a series of tests, which will then be automatically executed by the software, using different gases and conditions to test on the same sample.
The TotalPerm gas permeameter is the only instrument on the market that can test the permeability of three gases. TotalPerm is equipped with three sensors that can test oxygen, carbon dioxide, and water vapor. TotalPerm has the ability to test three different gas permeabilities on the same thin film sample surface, avoiding sample replacement.
The TotalPerm gas permeameter can test plastic films, single-layer or multi-layer barrier films, metal coatings or other surface coatings, multi-layer co extruded composite films, etc., especially in applications such as food, beverage, pharmaceutical, electronic packaging, etc. In addition to conducting penetration testing through thin films, TotalPerm can also be equipped with modular accessories for measuring various types of packaging containers, such as packaging bags, PET bottles, and packaging bags.
TotalPerm gas permeameter is an ideal solution for companies that produce or use barrier packaging, allowing for more functionality and higher performance on a single instrument to reduce overall costs. *The TotalPerm ExtraSolution? The software is very easy to use and operate.
TotalPerm is a superior choice for gas permeability testing, making experiments easier while reducing costs and saving space. You can use software to plan a series of tests, which will then be automatically executed by the software, using different gases and conditions to test on the same sample.
One TotalPerm can complete all the tests you need, so why buy three different instruments?
Calibration method for TotalPerm gas permeameter
1) After obtaining the standard film from the manufacturer, conduct a visual inspection on the sample. Ensure that there are no wrinkles, creases, pinholes, contamination, or other issues with the standard film;
2) Conduct the test according to the manufacturer's specified test conditions. For specific testing steps of the TotalPerm gas permeameter, please refer to the instruction manual;
3) Repeatedly experiment and calculate the arithmetic mean of multiple (≥ 3) reasonable air permeability tests;
4) If calibration is required, enter the arithmetic mean of the air permeability in step 3 in the "Test Current Value" section of the calibration interface. Enter the standard film experimental data provided by the manufacturer in the "Film Standard Value" section. Click the 'Calibration' button. Click the "OK" button to exit and complete the calibration.
5) After calibration, it is necessary to use the standard film provided by the manufacturer for verification.
Calibration process of TotalPerm gas permeameter
If the arithmetic mean of the standard membrane after multiple tests is within 10% of the data provided by the manufacturer, calibration is not required; On the contrary, calibration and verification should be carried out according to steps 4 and 5.