Welcome Customer !

Membership

Help

CYIPMI China Regional Service Office (Porous Measurement Int'l Group)
Custom manufacturer

Main Products:

instrumentb2b>Products

CYIPMI China Regional Service Office (Porous Measurement Int'l Group)

  • E-mail

  • Phone

  • Address

    Shanghai

Contact Now

Gas micro flow permeability analyzer MGP

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The gas micro flow permeability analyzer MGP can accurately measure the important permeability and gas permeability characteristics of porous materials, such as micro flow gas permeability analysis, Darcy Law, Gurley permeability analysis, and Frazier permeability analysis. Breathability refers to the time required for a certain amount of air to pass through a diaphragm under a certain pressure and measurement area. High breathability indicates long breathability time and poor permeability. Generally refers to materials with good breathability, but the air permeability is not too high.

Product Details

principle

♦ gas permeability

It is the volume of gas that passes through a unit area of the sample per unit time, determined by two factors: the barrier properties of the material itself and the pressure difference of the tested gas on both sides of the material. The gas permeability is related to the testing environment and the individual sample, while the gas permeability is only related to the testing sample itself and not to the environment. As for the gas permeability coefficient, it is only related to the material of the sample and is a property of the material itself.

Gas permeation rate (Q) is the volume of gas that passes through a unit area of a sample per unit time at a constant temperature and unit pressure difference, under stable permeation. The unit is cm3/m2•d•Pa。 There is a relationship between d and gas permeability, where d is the thickness of the material.Due to the difference in units between the two, it is necessary to unify the calculation units. For example, when the unit of gas permeability coefficient of a material is cm3•cm/cm2• s • Pa and the unit of gas permeability is cm3/m2When d • Pa is used, the difference in test time units introduced during the calculation process alone is 86400 times, and the difference in area units is 10000 times. Therefore, the national standard GB1038 provides 1.1574X10-9This coefficient is used for the unification of units.

specification

♦ Hardware part:

Transmittance range 1 x 10-101x 10-6(Microflow in cc/sec/m/torr)
pressurized gas Clean, dry, and non corrosive gases (N2, AIR...)
Test pressure range 0~500 PSI
Pressure conversion range 0~500 PSI
Pressure measurement and regulation control system fully automatic control
sample size 0.5' - 2.5' diameter
Sample testing types Thin film, paper, non-woven fabric, nanofiber membrane, PTFE membrane, etc
Range of traffic conversion 0.00001cc~10 cc/minute
Flow measurement system fully automatic control
sample slot

stainless steel material

Operation control Can choose between automatic testing and manual operation modes

♦ Software section:

  • Using USB connection and compatible with Windows 7/8/10 operating system
  • The testing mode has two modes: automatic testing mode and manual operation mode
  • Capwin operating software