- Phone
-
Address
Shanghai
CYIPMI China Regional Service Office (Porous Measurement Int'l Group)
Shanghai
characteristic
principle
The Frazier permeability tester is used to determine the gas permeability of porous solids. Let the gas flow of air (or other gases) pass through the test sample, and calculate the permeability data (in Darcy and other units) based on the stable flow rate under a fixed upstream and downstream gas flow pressure difference condition.

specification
| sample size | Diameter 5cm2~100cm2Other requirements can be designed separately |
| Sample clamp cover | Manual or pneumatic (180N pressure), other requirements can be designed separately |
| pressurized gas | Non corrosive gases such as nitrogen and air |
| Pressure gauge range | 0~100 psi |
| flow range | 0~200l/min (other requirements can be designed separately) |
| pressure difference control | 0.5 ± 0.01 inch watertorr |