Technical Indicators:
1、 Gas phase part
1.1 Program heating retention time repeatability(RSD):≤0.1%(Actual measurement 0.04%);
1.2 Retention time difference:0.005min;
1.3 Excluding column boxes6 independent heating zones, 2 injection ports, 2 detectors, 2 auxiliary heating zones, maximum temperature in the auxiliary heating zone: 400 ℃;
1.4 diversion/Non split (S/SL) capillary injection port
1.4.1最高使用温度:400℃;
1.4.2Pressure setting range:0~700kPa;
1.4.3Total flow setting range:0~500 mL/min(N2);
1.4.4Maximum diversion ratio:600:1;
1.4.5Voltage control mode: constant current, constant voltage, program boost, program boost, pulse pressure;
1.5 column box
1.5.1Internal dimensions:280mm×310mm×160mm;
1.5.2Cooling method: rear door opening+Variable frequency motor;
1.5.3Operating temperature: above room temperature4℃~450℃;
1.5.4Resolution of temperature setting value:±0.1℃;
1.5.5Maximum heating rate of column box:120℃/min;
1.5.6Maximum running time:999 minutes;
1.5.7Every change in room temperatureChange in column temperature at 1 ℃: < 0.01 ℃;
1.5.8Cooling time of column box: fromReduce 250 ℃ to 50 ℃ ≤ 4 minutes;
two、Mass spectrometry section
2.1 Ambient temperature15-35℃;
2.2 relative humidity≤75%;
2.3 Applicable power supply:220V(±10%),50Hz(±2%);
2.4 Ion source
2.4.1 Ion source material: Inert ceramic ion source, equipped with dual filaments; The maximum filament current is 400MA;
2.4.2 Sensitivity: EI Sensitivity: 1 pg OFN full scan mode S/N: 1000:1;
2.4.3 Ionization energy: 10-100eV;
2.4.4 Interface temperature: adjustable from 50 ℃ to 350 ℃, with an accuracy better than 0.1 ℃. The mass spectrometer is temperature controlled and does not occupy the auxiliary heating area of the gas chromatograph;
2.4.5 Ion source temperature: 50 ℃ -350 ℃;
2.5 Quality Analyzer
2.5.1 Quality analyzer: high-precision inert metal quadrupole with pre-set rod;
2.5.2 Quadrupole rod temperature: Quality stability can be ensured without temperature control;
2.5.3 Quality range: 1.5-1050 amu, software can set and detect;
2.5.4 Quality accuracy: 0.1u;
2.5.5 Quality axis stability: better than 0.1 amu/48 h;
2.5.6 Scanning methods: full scan (SCAN), selective ion monitoring (SIM), full scan, and selective ion synchronous monitoring;
2.5.7Select ion monitoring mode(SIM)canUnrestricted grouping;
2.6 Detector
2.6.1 Detector: Long life 13 level discontinuous dynode electron multiplier;
2.6.2 Maximum scanning speed: 20000 amu/s, with adjustable speed throughout the entire process;
2.6.3 Dynamic range: 106;
2.6.4 Peak area reproducibility; Automatic injection RSD ≤ 5%;
2.6.5 Detection limit (IDL):<10fg; Octafluoronaphthalene, EI
2.7 Vacuum system: Mechanical pump pumping speed 4m3/h. Configure high-performance turbo molecular pump with a pumping speed of 250 L/s;
2.8 Maximum allowable chromatographic column flow rate 5
Standard configuration:
GC138N gas chromatography host
●MS8300 mass spectrometrydetector
●DB-5MS capillary chromatography column
●Large capacity helium capture trap
Optional:
●AS9Liquid automatic sampler
●Thermal desorption apparatus
●Blow and captureceremony
●Fully automatic headspace sampler