- Phone
-
Address
101, Private No.1 Comprehensive Building, Intersection of Mingzhu Road and Taiping Road, Jianhu County, Jiangsu Province
Yancheng Tianyue Instrument Co., Ltd
101, Private No.1 Comprehensive Building, Intersection of Mingzhu Road and Taiping Road, Jianhu County, Jiangsu Province
GXH-3011A Portable Infrared CO Analyzer
1、 Overview
This instrument complies with the national standards of GB/T18204.2-2014 "Methods for hygiene inspection in public places - Part 2: Chemical pollutants", GBZ/T300.37-2017 "Determination of toxic substances in workplace air - Part 37: Carbon monoxide and carbon dioxide", and GB/T9801-1988 "Determination of carbon monoxide in air quality - Non dispersive infrared method"; Complies with the standard HJ/T44-1999 "Determination of Carbon Monoxide in Fixed Source Exhaust Gas by Non Dispersive Infrared Absorption Method".
This instrument complies with the national metrological verification regulations of JJG635-2011 "Infrared Gas Analyzer for Carbon Monoxide and Carbon Dioxide", and its main technical indicators meet the technical requirements of national secondary instruments. It can obtain a verification certificate from the Chinese Academy of Metrology (verification conclusion: qualified, this instrument meets the secondary technical requirements).
This instrument has a built-in zero adjustment filter and a six way valve for switching between zero adjustment and measurement, making it easy and flexible to operate.
This instrument can customize the measurement range.
2、 Main technical indicators
Measurement principle: Non dispersive infrared analysis method/Non dispersive infrared method (NDIR)
Sampling method: built-in pump suction
Measurement range: 0~50.0 × 10-6Or 0~200.0 × 10-6
Resolution: 0.1 × 10-6
Repeatability: ≤ 1% full scale
Zero drift: ≤± 2% at full scale/h
Span drift: ≤± 2% full scale/3h
Linear deviation: ≤± 2% at full scale
Preheating time: ≤ 30 minutes
Response time: t0~t90≤45S
Flow range: (0.5-1.0) L/min
Standard configuration: host, sampler, battery, charger, technical documents, portable case
Power supply: AC/DC dual-use, 220AVC (± 10%) or built-in rechargeable battery
Dimensions: 245 × 190 × 85 (mm)
Weight: ≤ 3kg