E-mail
testservice@126.com
-
Phone
18615268674
-
Address
No. 15 Lanxiang Road, Tianqiao District, Jinan City
Shandong Puchuang Industrial Technology Co., Ltd
testservice@126.com
18615268674
No. 15 Lanxiang Road, Tianqiao District, Jinan City


GB/T 21529 Electrolytic Water Vapor Transmittance Tester
Product Introduction
The WVTR-AE1 electrolytic water vapor permeability tester is suitable for measuring the water vapor permeability of high barrier materials such as plastic films, aluminum foil plastic films, waterproof materials, and metal foils. Expandable test bottles, bags, and other containers.
Test Principle
Using the principle of electrolytic sensor testing, nitrogen with stable relative humidity flows on one side of the film, while dry nitrogen flows on the other side of the film; Due to the presence of humidity gradient, water vapor will diffuse through the film from the high humidity side to the low humidity side; On the low humidity side, the water vapor passing through is carried to the sensor by the flowing dry nitrogen gas. When entering the sensor, an electrical signal of the same proportion is generated. By analyzing and calculating the electrical signal of the sensor, parameters such as the water vapor transmission rate of the sample can be obtained. For packaging containers, dry nitrogen gas flows inside the container while the outside of the container is in a high humidity state.
Technical Specifications
Measurement range (membrane) 0.01~40 g/(m2 • day) (standard)
0.1~1000 g/(m2 • day) (optional)
Sample quantity1
resolution0.001 g/(m2 •day)
sample sizeΦ108mm
Test area50cm2
sample thickness≤ 3mm
Temperature control range15 ℃~55 ℃ (resolution ± 0.01 ℃)
temperature control accuracy±0.1℃
Moisture control range0 ~ 100%RH
Moisture control accuracy± 1%RH
Carrier gas99.999% high-purity nitrogen (gas source to be provided by the user)
Carrier gas flow rate0~200ml/min (fully automatic control)
air supply pressure≥0.28MPa/40.6psi
interface dimensions1/8'
Correction methodStandard film calibration
host size350(L)×695(W)×410(H)mm
host weight60Kg
power supplyAC 220V 50Hz
reference standard
YBB 00092003、GBT 26253、ASTM F1249、ISO 15106-2、TAPPI T557、JIS K7129、ISO 15106-3、GB/T 21529、DIN 53122-2
GB/T 21529 Electrolytic Water Vapor Transmittance Tester