Fully automatic solvent residue analyzer

NegotiableUpdate on 08/02
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The EZ-VOC series fully automatic solvent residue analyzer is designed based on the principle of headspace gas chromatography technology, integrating automatic high-sensitivity headspace injection and EPC high-resolution capillary column gas chromatography with excellent performance. Therefore, it has the characteristics of fast, reliable, excellent reproducibility, and high analytical accuracy. Easy to operate, even non professional technicians can easily use it.

Product Details

EZ-VOCFully automatic solvent residue analyzerModel and application scope:

EZ-VOC-I: Suitable for solvent residue and odor analysis in packaging materials;
EZ-VOC-II: Suitable for solvent residue analysis in pharmaceuticals;
EZ-VOC-III: Suitable for analysis of alcohol residues in blood;
EZ-VOC-EO: Suitable for residual analysis of ethylene oxide in disposable medical products;
EZ-VOC-Flavor: It is applicable to the analysis of conventional flavor components of foods (such as beverages, beer, Baijiu, etc.);
EZ-VOC Beer: suitable for trace analysis of beer flavor components;
EZ-VOC-TOGA: Suitable for transformer oil analysis, in compliance with ASTM D3612 standard.
EZ-VOCFully automatic solvent residue analyzerMain technical indicators:
★ Detection limit for solvent residue: ppm to sub ppm level (depending on the actual sample situation)
★ Headspace sample pretreatment: 40 samples, simultaneously heating/oscillating 6 samples at a temperature of 150 ℃; The injection volume and injection speed can be adjusted arbitrarily
★ Gas chromatograph: large volume column box, unique double door design, dual power supply system, EPC electronic flow control
High power vortex blower device ensures rapid cooling and uniformity of column box temperature up to 0.1 ℃
FID detector, no dead volume structure, sensitivity 10-12G/s dynamic range: better than 108
Chromatography workstation: supports network data acquisition, automatic calibration, and automatic batch processing functions