Fully automatic high and low temperature cold shock test chamber

NegotiableUpdate on 01/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Installation and commissioning of fully automatic high and low temperature cold shock test box: Place the test box in a well ventilated, dry, and vibration free place. Connect the power supply of the test chamber and check if the electrical system of the equipment is functioning properly. Open the control panel of the test chamber and set the test parameters, such as temperature range, temperature change rate, test time, etc.
Product Details

Fully automatic high and low temperature cold shock test chamberThe structure:

Internal dimensions 80L: W400 × H500 × D400mm

Inner box material: stainless steel plate

Outer box material baked paint

Insulation material a. High temperature tank: Aluminum silicate insulation cotton

b. Low temperature tank: high-density PU foam

Single panel doors for the upper, middle, and lower compartments, open to the left

a. Embedded flat handle

b. Rear button: SUS # 304

c. Silicone foam strip

Sample placement rack a. Intermediate testing room

b. The load-bearing capacity shall not exceed 5KG

c. Stainless steel SUS # 304 inner box

Fin type heat dissipation tube shaped stainless steel electric heater for heating system

1. High temperature tank 6KW

2. Low temperature tank 3.5KW

Air circulation system 1. Two 1HP motors

2. Stainless steel extended axis

3. Multi blade fan (SIROCCOFAN)

4. Specially designed fan forced air circulation system

Inner box material: mirror stainless steel plate (SUS # 304)


Fully automatic high and low temperature cold shock test chamberUser manual:

1、 Equipment installation and debugging

Place the test chamber in a well ventilated, dry, and vibration free location.

Connect the power supply of the test chamber and check if the electrical system of the equipment is functioning properly.

Open the control panel of the test chamber and set the test parameters, such as temperature range, temperature change rate, test time, etc.

Perform device self inspection to ensure that all components of the device are functioning properly.

2、 Sample preparation

Clean the surface of the sample to ensure that there are no impurities such as dust, grease, etc.

According to the testing requirements, fix the sample to prevent displacement or damage during the testing process.

Label the samples so that they can be quickly found after the experiment is completed.

3、 Experimental operation

Place the sample in the test chamber and be careful not to block the air duct to avoid affecting the test results.

Close the door of the test chamber and ensure that it is well sealed.

Set test parameters, such as temperature range, temperature change rate, test time, etc.

Start the test chamber, observe the operation of the equipment, and ensure that the equipment is working properly.

During the experiment, the status of the sample can be observed through the observation window of the test chamber. If there is any abnormal situation, the experiment should be stopped immediately.

After the experiment is completed, open the door of the test chamber and take out the sample, taking care to avoid burns.

4、 Equipment maintenance and upkeep

Regularly clean the interior of the test chamber to remove dust, grease, and other impurities.

Check the electrical system of the equipment to ensure that the wires, plugs, and other components are intact and undamaged.

Check the refrigeration system of the equipment, such as compressors, condensers, evaporators, and other components, to ensure their normal operation.

Check the heating system of the equipment, such as heaters, fans, and other components, to ensure that they are working properly.

Regularly calibrate the equipment to ensure the accuracy of the test results.

5、 Safety precautions

Operators should possess certain electrical knowledge and operational skills, and non professionals are strictly prohibited from operating equipment.

During the operation of the equipment, it is strictly prohibited to touch the high-temperature parts of the equipment to avoid burns.

When repairing equipment, the power should be cut off first to ensure that the equipment is in a shutdown state.

Set up clear safety warning signs around the equipment to remind other personnel to pay attention to safety.

VIFully automatic high and low temperature cold shock test chamberfunctionAbnormal judgment and handling:

form moreover

original because

place reason square law

End screen display

The experimental period has ended.

Defrost screen display

The defrosting frequency has been reached.

Defrost until the screen disappears.

Display of over temperature abnormal screen

1. The over temperature protection switch of the high and low temperature preheating tank is set too low.

1. Please check if the over temperature protection switch is set within the required range:

High temperature preheating zone: 300 ° C

Low temperature pre cooling zone: 100 ° C

2. Overtemperature protection switch malfunction

1. New products need to be replaced.

3. SSR fault penetration

1. New products need to be replaced.

The first section of the compressor abnormal screen display

1. The overload protection device of the compressor has tripped.

1. Press back the reply button of the overload protection switch.

2. The high and low pressure protection switch of the compressor has tripped.

1. Wait for a few minutes before manually resetting. If you jump off again, please notify our company

3. Compressor overheating

1. If the ambient temperature is too high, please control the ambient temperature within 15 ℃~28 ℃.

2. Check if the cooling water is overheated. If it is overheated, check if the cooling fan of the water tower is running. If it is not running, update the cooling fan.

Display of abnormal screen for the second stage compressor

1. The overload protection device of the compressor has tripped.

1. Press back the cover button of the overload protection switch.

2. The high and low pressure protection switch of the compressor has tripped.

1. Wait for a few minutes before manually resetting. If you jump off again, please notify our company

3. Compressor overheating

1. If the ambient temperature is too high, please control the ambient temperature within 15 ℃~28 ℃.

2. Check if the cooling water is overheated. If it is overheated, check if the cooling fan of the water tower is running. If it is not running, update the cooling fan.