10Fully automatic gas sampler
Can perform batch injection of 10 airbags;
Independently developed software control;
Clear display of the status of the instrument system;
Capable of controlling chromatographic instruments for sequential injection;
Capable of automatic sample injection control through an analyzer to improve analysis efficiency;
The injection port is a pagoda joint, which is convenient for connecting the airbag hose and can be customized according to user needs;
Use a vacuum diaphragm pump for injection to ensure sample reproducibility;
Connect the sample to the gas injection valve of the chromatograph using a passivated stainless steel tube with an outer diameter of 1/16 ".
10Fully automatic gas sampler
Technical parameters:
Number of injection ports: 10;
External interfaces: serial port, touch EVENT、BCD;
Power: 100W;
Net weight: 10kg;
Connecting pipeline: 1/16 inch passivated stainless steel pipeline;
Power supply: 220V, 50Hz;
Dimensions: 400 * 480 * 630mm.
Parameter Description
Xun loop count: The number of Xun loops used for sample injection analysis
Xun ring waiting time: the duration of waiting before the injector injects the sample
Injection time: The duration from the intake of air to the start of data collection by the gas chromatograph
Injection valve action: duration of air intake into the bladder
Chromatography startup: the duration of sending a signal to the chromatograph
Injection pump action: duration of air intake into the bladder and duration of negative pressure discharge from the sampler pipeline
Sample analysis duration: The duration of chromatographic analysis.
Electromagnetic valve testing: Without analysis, test the signals of the solenoid valve, pump, and chromatograph.
CH01-10: Electromagnetic valve 1-10
CH16: Chromatograph signal
CH16: Pump
Click on the blank space in the upper left corner: Save overall parameters
Click stop during analysis, the stop button turns purple, click cancel stop again.