Fully automatic flame photometerProduct features
1. Direct reading of potassium and sodium element concentrations;
2. Menu based keyboard operation;
3. Automatic calculation of correlation coefficient;
4. Pre select the flame size;
5. Flameout protection device, safe and reliable;
6. Direct printing of data requires an external printer;
7. Connect to a computer for data processing.

Fully automatic flame photometerThe completeness of the set
1) 1 instrument (host)
2) 1 air compressor
3) 1 set of accessory spare parts
Normal working conditions
1. Environmental temperature (10-35) ℃.
2. Relative humidity ≤ 85%.
3. The power supply voltage is (220 ± 22) V, the frequency is (50 ± 1) Hz, and it has good grounding.
4. The rated power is 30W.
5. The product should be placed horizontally on a non vibrating workbench, and there should be no shaking during operation.
6. The testing site should have good ventilation and should not be directly exposed to strong light. There should be no interference from strong magnetic fields, electric fields, or vibration sources around the product, and no strong airflow.
7. There should be no flammable, explosive, or corrosive gases on site when using liquefied gas cylinders for the product, and fire extinguishing facilities should be prepared.
technical indicators
Stability: Continuous injection with standard solution, relative maximum change in instrument reading within 15 seconds ≤ 3%
Measure once per minute for a total of 6 measurements, with a relative maximum change in instrument readings of ≤ 15%
Repeatability: ≤ 3%
Linear error: K: ≤ 0.005mol/L (0.0100-0.0800) mmol/L;
Na:≤0.03mmol/L(0.0500~0.400)mmol/L
Detection limit: K: Detection limit ≤ 0.004mmol/L;
Na: Detection limit ≤ 0.008mmol/L
Transmittance characteristics of filter: Potassium filter: absolute value of peak wavelength error ≤ 7nm;
Sodium filter: absolute value of peak wavelength error ≤ 5nm
Potassium filter: half width ≤ 15nm;
Sodium filter: Half width ≤ 15nm
Potassium filter: background transmittance ≤ 0.5%;
Sodium filter: background transmittance ≤ 0.5%
Response time:<8s
Sample suction and spraying volume: < 6mL/min
Display mode: LCD screen display
Spectral splitting method: interference filter
Optoelectronic conversion element: silicon photovoltaic cell
Dimensions l × b × h: 400 × 250 × 500 (mm × mm × mm)
Mass (weight) kg: 8.0
Test range: K: 0-100ug/ml
Na:0~160ug/ml
Fully automatic flame photometerPrecautions for Operation
1) Keep the instrument room clean and ventilated;
2) Use a stable 220V power supply voltage;
3) Gas and auxiliary gas (air) must be dry and pure;
4) During the operation, the body should not approach or touch the combustion chamber and chimney cover, nor should it look down from top to bottom;
5) After successful ignition, wait for 10-20 minutes before calibration, otherwise the calibration value will be inaccurate;
6) The combustion gas of the instrument should be turned on at the latest, and the first step after use is to turn off the combustion gas;
7) After preparing the standard solution, it should be stored in a polyethylene plastic bottle, and the discharged waste liquid should be collected and treated centrally;
8) Keep the atomization chamber and combustion head clean and maintained. If high salt concentration samples are tested, the time for spraying distilled water should be appropriately extended;
9) The sample to be tested must be clear, and attention should be paid to the liquid level during operation to ensure that the injection tube only absorbs the upper layer of solution.
10) To avoid the impact of previous tests on the testing, enter the "Clear" menu in the initial menu and clear all the data from the previous "Calibration" and "Testing".
11) After use, it must be cleaned with alcohol (distilled water) until the pressure gauge pointer reaches zero.