The formaldehyde testing chamber is also suitable for detecting the emission of pollutants from products such as automotive parts and plastic tracks. This product has obtained the national invention license, license number 2015102462743.

Product Details

Product Introduction

Formaldehyde testing chamberMainly used for detecting the emission of pollutants, including formaldehyde and VOCs, from materials and products such as small furniture, building panels, paint coatings, and small equipment. The same applies to detecting the emission of pollutants from products such as automotive parts and plastic tracks. This product has obtained the national invention license, license number 2015102462743.

1. Extremely low background concentration;

2. Temperature and humidity display and control;

3. Ensure uniform temperature to prevent condensation;

4. The wind speed of the mixing fan can be adjusted to meet the requirements of the testing space wind speed;

5. The amount of clean air can be precisely adjusted;

6. Extremely low work noise creates a quiet workspace;

7. Parameters such as temperature and humidity, clean air supply volume, and pressure difference can be downloaded remotely;

8. Computer monitors the operation of multiple devices.


Formaldehyde testing chamberThe testing principle:

The process of testing pollutant release in an environmental chamber can be summarized as follows: placing materials and products or their samples in the testing chamber, adjusting the environmental conditions such as temperature, relative humidity, and air replacement rate in the chamber as needed; Sampling at the exhaust outlet or inside the environmental chamber; Using appropriate methods to analyze gas samples, identify the main pollutants and their concentration changes over time; Then determine the release rate and release characteristics of the material and product based on the measured concentration.


Design reference standards for environmental cabin:

ASTM D 5116-97 Standard Guide for Small-Scale Environmental Chamber Determinations 0f Organic Emissions From Indoor Materials/Products

ASTM D 6330-98 Standard Practice for Determination of Volatile Organic Compounds (Excluding Formaldehyde) Emissions from Wood-Based Panels Using Small Environmental Chambers Under Defined Test Conditions

ASTM D 6670-01 Standard Practice for Full-Scale Chamber Determination of Volatile Organic Emissions from Indoor Materials/Products


E 1333—96 Standard Test Method for Determining Formaldehyde Concentrations in Air and Emission Rates from Wood Products Using a Large Chamber

ASTM D 6007-02 Standard Test Method for Determining Formaldehyde Concentration in Air from Wood Products Using a Small Scale Chamber

ENV 13419-1 Building products-Determination of the emission of volatile organic compounds-Part 1:Emission test chamber method

ENV 717-1 Test on the Formaldehyde emission , “Wood-Based Panels Determination of Formaldehyde Release-Part 1:Formaldehyde Emission by the Chamber Method”


Technical Specifications:

Project

technical indicators

temperature

Adjustment range: (15-70) ℃

Adjustment accuracy: ± 0.3 ℃

relative humidity

Adjustment range: (30-80)% RH

Adjustment accuracy: ± 3% R.H

Air exchange rate

Adjustment range: 0.2 m3~2.0 m3/h

(Ventilation rate 0.2-2.0 times/hour adjustment accuracy: Ventilation rate ± 3%)

Face Velocity

Maintain a horizontal air velocity of (0.1-0.3) m/s

Test cabin size (inner container)

800*1000*1250mm, Deviation not exceeding 0.02m3

SUS304 mirror stainless steel plate is used for overall welding, with polished welding points and surrounding areas without deformation

Relative pressure difference inside the cabin compared to the outside

15 ± 5Pa, adjustable

sealing performance

Static leakage rate: Under a relative pressure of 1KPa, the leakage rate is ≤ 1 × 10-3 m3/min

insulated box

Polyurethane insulation board with a thickness of 80mm, outer wall made of steel plate, sprayed; Inner wall: 304 stainless steel plate

Background concentration

Formaldehyde background concentration<0.006mg/m3, TVOC background concentration<0.02mg/m3. Single VOC background concentration<0.005mg/m3

measurement and control system


Real time monitoring and recording of parameters such as temperature, humidity, and clean air volume inside the cabin. Various parameters can be easily set on the control screen. It has functions such as dynamic data display and alarm.

structural design

By adopting the capillary network temperature control technology with independent intellectual property rights, the temperature in the testing chamber is more uniform and the temperature fluctuations are very small.

noise

The working noise is below 60dBA

Test cabin exterior dimensions

Width 1200mm x Height 1900mm x Length (Depth) 2000mm

power

Energy saving design, with an average power not exceeding 1.5KW during equipment operation.

Power requirements

220V

environmental conditions

1) Temperature: 10 ℃~35 ℃, stable ambient temperature

2) Relative humidity: ≤ 85%

3) Air pressure: 86kPa~106kPa

4) There is no pollution source around


Note: Various sizes of temperature controlled humidity chambers can also be customized according to customer requirements.