E-mail
hytdtec@163.com
-
Phone
17694800211,15811077873
-
Address
Room 6208A, Kangde Building, No.17 Shangdi 6th Street, Haidian District, Beijing
Beijing Hongyuan Tongda Technology Co., Ltd
hytdtec@163.com
17694800211,15811077873
Room 6208A, Kangde Building, No.17 Shangdi 6th Street, Haidian District, Beijing
Product Introduction
Formaldehyde testing chamberMainly used for detecting the emission of pollutants, including formaldehyde and VOCs, from materials and products such as small furniture, building panels, paint coatings, and small equipment. The same applies to detecting the emission of pollutants from products such as automotive parts and plastic tracks. This product has obtained the national invention license, license number 2015102462743.
1. Extremely low background concentration;
2. Temperature and humidity display and control;
3. Ensure uniform temperature to prevent condensation;
4. The wind speed of the mixing fan can be adjusted to meet the requirements of the testing space wind speed;
5. The amount of clean air can be precisely adjusted;
6. Extremely low work noise creates a quiet workspace;
7. Parameters such as temperature and humidity, clean air supply volume, and pressure difference can be downloaded remotely;
8. Computer monitors the operation of multiple devices.
Formaldehyde testing chamberThe testing principle:
The process of testing pollutant release in an environmental chamber can be summarized as follows: placing materials and products or their samples in the testing chamber, adjusting the environmental conditions such as temperature, relative humidity, and air replacement rate in the chamber as needed; Sampling at the exhaust outlet or inside the environmental chamber; Using appropriate methods to analyze gas samples, identify the main pollutants and their concentration changes over time; Then determine the release rate and release characteristics of the material and product based on the measured concentration.
Design reference standards for environmental cabin:
ASTM D 5116-97 Standard Guide for Small-Scale Environmental Chamber Determinations 0f Organic Emissions From Indoor Materials/Products
ASTM D 6330-98 Standard Practice for Determination of Volatile Organic Compounds (Excluding Formaldehyde) Emissions from Wood-Based Panels Using Small Environmental Chambers Under Defined Test Conditions
ASTM D 6670-01 Standard Practice for Full-Scale Chamber Determination of Volatile Organic Emissions from Indoor Materials/Products
E 1333—96 Standard Test Method for Determining Formaldehyde Concentrations in Air and Emission Rates from Wood Products Using a Large Chamber
ASTM D 6007-02 Standard Test Method for Determining Formaldehyde Concentration in Air from Wood Products Using a Small Scale Chamber
ENV 13419-1 Building products-Determination of the emission of volatile organic compounds-Part 1:Emission test chamber method
ENV 717-1 Test on the Formaldehyde emission , “Wood-Based Panels Determination of Formaldehyde Release-Part 1:Formaldehyde Emission by the Chamber Method”
Technical Specifications:
Project |
technical indicators |
temperature |
Adjustment range: (15-70) ℃ Adjustment accuracy: ± 0.3 ℃ |
relative humidity |
Adjustment range: (30-80)% RH Adjustment accuracy: ± 3% R.H |
Air exchange rate |
Adjustment range: 0.2 m3~2.0 m3/h (Ventilation rate 0.2-2.0 times/hour adjustment accuracy: Ventilation rate ± 3%) |
Face Velocity |
Maintain a horizontal air velocity of (0.1-0.3) m/s |
Test cabin size (inner container) |
800*1000*1250mm, Deviation not exceeding 0.02m3 SUS304 mirror stainless steel plate is used for overall welding, with polished welding points and surrounding areas without deformation |
Relative pressure difference inside the cabin compared to the outside |
15 ± 5Pa, adjustable |
sealing performance |
Static leakage rate: Under a relative pressure of 1KPa, the leakage rate is ≤ 1 × 10-3 m3/min |
insulated box |
Polyurethane insulation board with a thickness of 80mm, outer wall made of steel plate, sprayed; Inner wall: 304 stainless steel plate |
Background concentration |
Formaldehyde background concentration<0.006mg/m3, TVOC background concentration<0.02mg/m3. Single VOC background concentration<0.005mg/m3 |
|
measurement and control system |
Real time monitoring and recording of parameters such as temperature, humidity, and clean air volume inside the cabin. Various parameters can be easily set on the control screen. It has functions such as dynamic data display and alarm. |
structural design |
By adopting the capillary network temperature control technology with independent intellectual property rights, the temperature in the testing chamber is more uniform and the temperature fluctuations are very small. |
noise |
The working noise is below 60dBA |
Test cabin exterior dimensions |
Width 1200mm x Height 1900mm x Length (Depth) 2000mm |
power |
Energy saving design, with an average power not exceeding 1.5KW during equipment operation. |
Power requirements |
220V |
environmental conditions |
1) Temperature: 10 ℃~35 ℃, stable ambient temperature 2) Relative humidity: ≤ 85% 3) Air pressure: 86kPa~106kPa 4) There is no pollution source around |
Note: Various sizes of temperature controlled humidity chambers can also be customized according to customer requirements.
Control screen interface