1YJ-2102CFluorescence full wavelength enzyme-linked immunosorbent assay (ELISA) readerFeatures
1. Adopting imported grating and PMT design, with absorption light, fluorescence, and chemiluminescence detection functions
2. HavingAbsorption spectrum scanningFluorescence excitation and emission spectrum scanning
3. The wavelength range of the monochromator is 190~1000nm, and 1/5/10nm/step can be set
4. The light source adopts imported high-energy scintillation xenon lamp, with a long service life of up to 109 times
5. Equipped with a temperature controlled incubation system, with a range of RT+5~50 ℃ and a temperature control accuracy of ± 0.5 ℃ at 37 ℃
6. Equipped with microporous plate shaking and mixing function, with optional strength levels of high, medium, and low, and a 24-hour adjustable time
IIYJ-2102CFluorescence full wavelength enzyme-linked immunosorbent assay (ELISA) readerMain technical parameters
Absorption light (grating type)
1. Light source: high-energy flashing xenon lamp
2. Grating: Imported grating
3. Detection system: 1 detection channel+1 reference channel, PD detector
4. Wavelength range: 190 1000nm, 1/5/10nm/step
5. Reading range: 0.000-4.000Abs (@ 450nm)
6. Linear range: 0.000-3.000Abs (@ 450nm)
Fluorescence (grating type)
1. Light source: high-energy flashing xenon lamp
2. Detector: High performance photomultiplier tube (PMT)
3. Excitation wavelength range: 200 1000nm, resolution 1nm, supports excitation spectrum scanning
4. Emission wavelength range: 250 850nm, resolution 1nm, supports emission spectrum scanning
5. Sensitivity: ≤ 0.5 fmol (FITC)
6. Detection limit: ≥ 6 orders of magnitude
7. Repeatability of absorbance: CV ≤ 0.5%
8. Absorption stability: ≤± 0.005 Abs
9. Error of absorbance reading ≤± 0.01 Abs
10. Wavelength accuracy: ≤± 1nm
11. Wavelength repeatability: ≤± 0.5nm
chemiluminescence
1. Detector: Photon Counting Photomultiplier Tube (PMT)
2. Wavelength range: 270 850nm
3. Sensitivity: ≤ 5mmol/well
4. Inter hole interference:<0.2%
5. Syringe: optional
Other parameters and functions
1. Display: 10.1 LCD touch screen
2. Operation: Single machine operation, can also be connected to a computer for operation
3. Interface: COM+USB (communication, printing, USB drive)
4. Power input: Using an adapter to input 160W (DC24V 6.67A),
Network power supply AC100V-240V 50/60Hz, full range support
5. Product Quality: Approximately 30kg
6. Dimensions (length * width * height): approximately 550mm * 440mm * 360mm