FY2070 Coating Thickness Gauge

NegotiableUpdate on 11/20
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Introduction: The FY2070 coating thickness gauge adopts a multifunctional host, integrating magnetic and eddy current thickness measurement methods. The magnetic method is applied to measure the thickness of galvanized, copper, chromium and other coatings on steel, as well as spray coatings, anti-corrosion coatings, etc. It requires the use of an F-type probe eddy current method for measuring the thickness of oxide coatings on aluminum surfaces, copper or aluminum surface spray coatings, etc. It requires the use of an N-type probe

Product Details

FY2070Coating thickness gaugeFeatures:
Six types of probes (F400, F1, F1/90 °) can be used F10、N1、CN02);
Adopting the industry's latest high-frequency CPU and advanced integrated circuit design,The measurement response speed is faster than that of imported thickness gaugesImprove work efficiency.
The all metal aluminum brushed shell structure is more sturdy and durable, with extremely high measurement stability, and can be used for a long time without calibration.
Two methods, single point calibration and two-point calibration, can be used to calibrate the instrument, and the basic calibration method can be used to correct the systematic errors of the measuring head;
Storage function: capable of storing 20 sets of measurement data (with an additional 50 measurement values per set);
There is a high brightness backlight display, which is convenient for use in dimly lit environments. There is a buzzing sound during the operation process;
It has an error prompt function, which provides error prompts through screen display or beeping sound;
There are two shutdown modes: manual shutdown mode and automatic shutdown mode;
The low-power design concept relies on stable IC integrated circuits, and AA size ordinary batteries can standby for about 200 hours, making it easy to carry and suitable for field operations
FY2070Technical parameters of coating thickness gauge:

Probe model
F1
N1
F10
F400
working principle
magnetic induction
eddy current
magnetic induction
magnetic induction
Measurement range(um
0-1250
0-1250
0~10000
0-400
Low resolution limit(um)
0.1
0.1
1
0.1
indication error
one-point calibration(um)
±[3%H+1]
±[3%H+1.5]
±(3%H+5)
±[3%H+0.5]
Two-point calibration(um)
±[(13)%H+1]
±[(13)%H+1.5]
±[(1~3)%H+5]
±(1~3) %H+0.5
measurement conditions
*Small curvature radius(mm)
convex1.5
3
10
1
Matrix * small area diameter(mm)
F7
F5
40
F3
*Small critical thickness(mm)
0.5
0.3
2
0.2


Note:H­—Measure thickness value

Standard configuration:
hostStandard film(1Set)MatrixChargerTheF1orN1probe
Optional attachments:
Probe for other purposesF400(0-400um)TheF10(0-10000um)TheF1/90°(Right angle probe)TheCNO 2(Copper foil specific)

Beijing Times Yuanfeng Technology Co., Ltd

electricity Words:

biography Really: