Welcome Customer !

Membership

Help

Anhui Donglin Automation Instrument Complete Set Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Anhui Donglin Automation Instrument Complete Set Co., Ltd

  • E-mail

    1368553620@qq.com

  • Phone

    13625506311

  • Address

    Tianchang City, Chuzhou City, Anhui Province

Contact Now

End face thermistor WZPM-001Q

NegotiableUpdate on 07/23
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

End face thermal resistor WZPM-001Q is usually used in conjunction with display instruments, recording instruments, and electronic regulators by Anhui Donglin manufacturer. It can directly measure the surface temperature of liquids, vapors, gases, and solids in the range of -200 ℃ to 420 ℃ during various production processes. Due to its certain service life, it is inevitable for thermal resistors to malfunction, so it is necessary to master the fault handling methods.

Product Details

End face thermistor WZPM-001Q

name Temperature measurement range ℃ Allowable deviation △ t ℃ Graduation
End face copper resistor -50-100 ±(0.3+6.0×10-3|t|) Cu50 (G) or Cu100
End face platinum resistor -150-200 B-level: ± (0.30+0.005 | t |) Pt100

Measurement range and tolerance (unit: mm)
Name Temperature measurement range ℃ Allowable deviation △ t ℃ Division number
End face copper resistance -50-100 ± (0.3+6.0 × 10-3 | t |) Cu50 (G) or Cu100
End face platinum resistance -150-200 Class B: ± (0.30+0.005 | t |) Pt100

End face thermistor WZPM-001Q
Model number
Graduation
structure characteristics
Temperature measurement range ℃
Material of protective tube
thermal response time
specifications
nominal pressure
External wiring lead length L (mm)
Diameter d (mm)
M( mm)
WZCM-201

Cu50 Cu100

Fixed embedding or threaded connection -50-100 Copper T2-Y <15 Φ 6 M8x0 .75 Atmospheric pressure

500
1000
1500
2000
2500

WZPM-201 Pt100 -150-200 <15 Φ 8.7 M10x1
WZPM-201B
Note: WZPM-201B adopts the imported component WZPM-018, with an accuracy level of A.