Product Introduction
It was developed by our companyOxygenSP1102The oxygen analyzer adopts a fuel cell sensor, and the core components are an electrochemical trace oxygen sensor imported from the United States and a high-performance sensor imported from FranceARMWhen the measured gas passes through the sensor, the processor converts the oxygen content in the measured sample gas into an electrical signal, which is amplified by the circuit and then processed by the microcontroller to display the oxygen concentration value, forming a standard signal output. By combining advanced production and inspection processes, this machine has excellent performance. It has been widely used in industries such as energy and chemical engineering, power electronics, etc.
Technical Specifications
Product Model:OxygenSP1102
Product Name: Trace Oxygen Analyzer
Analyzing gases: trace amounts of oxygen in mixed gases
Analysis principle: electrochemistry
Service life:>2 Year (used under normal conditions)
Product Warranty:1 year
Analysis range:0~10PPM|100PPM|1000PPM
Analysis resolution:0.01PPM
Measurement accuracy: ≤±2%F.S
Response time:T90<60 second
Flow range:0.3~3L/M
Usage of traffic:2±0.1 L/M
Import and Export: External Φ6 stainless steel pipe
Environmental temperature and humidity:-20℃∽50℃ ≦90%RH
Digital signal:RS232/RS485(Modbus)
Analog signal:4∽20mA
Whole machine power consumption: ≤3W
Working voltage:AC100V~240V 50/60Hz
Shell material: Aluminum alloy
Sample gas temperature:-10~40℃
Sample gas pressure:20~150Kpapressure protection&compensation
Standard external dimensions:
Specification and dimensions: 3U standard -483mm × 137mm × 350mm (W × H × D)
4U standard -483mm × 179mm × 380mm (W × H × D)
Hole size: 3U standard -445mm × 135mm (W × H)
4U standard -445mm × 177mm (W × H)
Whole machine weight: approximately5KG
Built in vacuum pump: optional product features
Ordering instructions:
In order to ensure the accuracy and suitability of the analytical instrument you have ordered, please specify the following:
1.Instrument range, signal output form, and other requirements for the instrument
2.Range of changes in the concentration of the measured gas
3.Background gas composition and concentration
4.Other required accessories (such as pre-processing, computer, recorder, etc.)