The ET-3722 fully automatic closed flash point tester is used to determine the closed flash point value of petroleum products and is widely used in the railway, aviation, power, petroleum industries, and scientific research. Standard support: GB/T261 detection, ignition, alarm, cooling, printing. The entire experimental process is automatically completed with platinum heating wire (standard) and gas ignition method (optional). High power heating technology is adopted, with high heating efficiency and automatic adjustment of heating curve

Product Details

product overview

The ET-3722 fully automatic closed flash point tester is designed and manufactured in accordance with the requirements specified in the industry standard GB/T261 "Determination of Flash Point of Petroleum Products (Closed Cup Method)" of the People's Republic of China.
Applicable to the method specified in this standard for determining the flash point of petroleum products using a closed cup method, which is the lowest temperature at which a mixture of steam and air heated under specified conditions comes into contact with a flame and ignites.

Product characteristics

1. Adopting a color LCD large screen display, a fully Chinese human-machine dialogue interface, an unmarked keyboard, and a prompt menu guided input for parameters such as temperature, sample number, atmospheric pressure, and test date that can be preset.
2. Simulate and track the function curve of temperature rise and test time, with Chinese misoperation software prompt modification function and parameter prompt functions such as test date and test time.
3. Equipped with a standard RS-323485 computer interface, the lower computer stores 100 sets of historical data, which can be connected to the computer to store data in large capacity and can be saved for a long time. When transmitting data, the upper computer can modify the parameters of the lower computer in real time.
4. Automatically calibrate the influence of atmospheric pressure on the experiment and calculate the correction value.
5. Differential detection, automatic correction of system deviation.
6. Opening, ignition, detection, and printing data are automatically completed, and the test arm is automatically raised and lowered.
7. Electronic ignition, forced air cooling.

Other name