ELSD evaporative light detector

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

▲ Anti corrosion protection treatment for drift tube and flow path system. The special process coating of drift tube has self-cleaning function and can be cleaned free. $r $n ▲ Low temperature diversion operation mode, low-temperature atomization and evaporation, can also achieve high sensitivity to thermally unstable and semi volatile compounds. Equipped with a full-color touch screen, it has temperature, flow, and pressure alarm protection functions, and can monitor the working status of the instrument in real time.

Product Details

ELSD evaporative light detector

▲ Anti corrosion protection treatment for drift tube and flow path system. The special process coating of drift tube has self-cleaning function and can be cleaned free.

▲ Low temperature diversion operation mode, low-temperature atomization and evaporation, can also achieve high sensitivity to thermally unstable and semi volatile compounds.

Equipped with a full-color touch screen, it has temperature, flow, and pressure alarm protection functions, and can monitor the working status of the instrument in real time.
Order Number HP-ELSDUM5800
power supply 220V,50Hz
volume 460×330×220 (mm)
environmental conditions Room temperature 15-35 ℃, relative humidity<80%
Evaporation temperature range Room temperature -130 ℃
Temperature regulation increment 1℃
Temperature control accuracy ±1℃
gas requirements Nitrogen or air
Gas input pressure range 2 bar -5 bar
Gas pressure detection accuracy 0.01bar
Gas flow range 1 L/min - 4 L/min
Gas flow accuracy ≤ 1% or 0.02 L/min
Liquid flow range 10 μL/min - 3 mL/min
light source detection 650 nm semiconductor laser
detection element photomultiplier tube
baseline noise <0.05 mV
(1.0 mL/min 100% methanol, evaporation temperature 35.C, gas flow rate 3.0 L/min)
baseline drift <1.0 mV/30 min
(1.0 mL/min 100% methanol, evaporation temperature 35.C, gas flow rate 3.0 L/min)
Minimum detectable concentration <5.00-6 g/mL cholesterol solution in methanol
(1.0 mL/min methanol, evaporation temperature 35.C, gas flow rate 3.0 L/min)
Typical quantitative range 0.1 μ g~30 μ g (glucose)
Qualitative repeatability (6 measurements)
RSD6 ≤ 0.5 %
Qualitative repeatability (6 measurements) RSD6 ≤ 0.5%
Quantitative repeatability (6 measurements)
RSD6 ≤ 3.0 %
Quantitative repeatability (6 measurements) RSD6 ≤ 3.0%
signal output 0~1VDC
computer interface RS232