EKA-200 Particle Size TesterTechnical Specifications:
● Execution standard: GB/T 18854-2002 (ISO11171-1999)
● Light source: semiconductor laser
● Particle size range: 1 μ m to 400 μ m
Sensitivity: 1 μ m (ISO4402) or 4 μ m (c) (GB/T18854, ISO11171)
● Detection channels: 16 channels with any particle size set within the range of 1 μ m to 100 μ m or 4 μ m (C) to 70 μ m (C)
Sampling method: bottle type
Sampling volume: 0.2-100ml, with an interval of 0.1ml
Sampling volume accuracy: better than ± 0.5%
Detection speed: 5mL/min~80mL/min
Cleaning speed: 5mL/min~80mL/min
Maximum vacuum in the sampling chamber: 0.08MPa
Maximum pressure of sampling chamber: 1MPa
Test sample viscosity: up to 650cSt
Test sample temperature: 0 ℃~80 ℃
Working temperature: 0 ℃~60 ℃
Resolution: better than 10% (GBT18854-2002, ISO11171)
● Overlap error limit: 12000~24000 particles/mL depending on the selected sensor
● Repeatability: RSD<2%
Technical accuracy: ± 10%
● Output method of test results: screen display; Built in printer for printing; Output to external computer
Power supply: 100-245V, 49-62Hz,
Power: ≤ 200W
● Dimensions: Host 360 * 370 * 720mm Pneumatic pump 330 * 266 * 255mm
EKA-200 Particle Size TesterFeatures:
Adopting the principle of light blocking method, it has the advantages of fast detection speed, strong anti-interference ability, high accuracy, and good repeatability.
High precision sensors ensure high resolution and accuracy.
The precision syringe type sampling system achieves constant sampling speed and precise control of sampling volume.
Positive/negative pressure sampling chamber device, realizing sample degassing and high viscosity sample detection.
5.6-inch large screen color LCD touch screen, graphical menu display, touch operation, simple and convenient.
Built in particle pollution level standards such as GB/T14039, ISO4406, NAS1638, SAE749D, GJB420A, GJB420A, and two sets of user-defined particle size channel settings; Built in standards according to user requirements.
Up to 9900 channels with arbitrarily set particle size for easy particle size analysis.
It can simultaneously store four calibration curves (latex ball calibration curve, ACFTD calibration curve, ISOMTD calibration curve, ROST calibration curve), and can be easily switched to reduce conversion errors.
● The detection data storage function facilitates the archiving, retrieval, and analysis of detection data.
Equipped with a built-in printer, it can directly print out inspection reports.
Built in Chinese input method to achieve Chinese annotation of detection reports.
RS232 interface and powerful data software enable external computer control of instruments and processing of detection data.