DMT143L online dew point meterBriefly describe
The online dew point meter adopts Vaisala(Vaisala)DRYCAP®sensorDMT143LDew point transmitter is a compact dew point measuring instrument that can be directly installed on50 bar(725 psia)In the pressure system. vaisala(Vaisala)DRYCAP®Sensor technology has long-term high-performance performance.
This sensor is resistant to humid environments, thereforeDMT143LThe transmitter can perform very well in industrial applications with water splashes, including pipeline condensation during process system failures or start-up phases. The sensor is also highly resistant to particulate impurities, oil and gas, and most chemical gases, and is not sensitive to changes in flow rate.
DMT143LOnline Dew Point Meterfeature/benefit
1A compact dew point transmitter suitable for small industrial dryer applications
2Vaisala with automatic calibration function(Vaisala)DRYCAP®Sensor Technology
3The calibration cycle is as long as once every two years
4Dew point measurement range-80... +20ºC(-76...+140°F)
5Accuracy ±2 ºC (±3.6 ºF)
6Anti condensation and dew condensation
7Fast response time
8, with Vaisala(Vaisala)Handheld dew point meterDM70compatible
9Tracing back to the sourceNISTCalibration of(Including calibration certificate)
10ThroughRS485User port, convenient and fast maintenance and data transmission
11Excessive dew pointLEDLamp alarm
DMT143LOnline Dew Point Metertechnical indicators
measurement parameters
dew point temperature
measurement range(typical)
-80 . .. +20°C (-76 ... +140 °F) Td
Analog output range
option1 -80 . .. +20°C (-112 ... +68 °F) Td
option2 -80 . .. +20°C (-112 ... +68°F) TdDew point at atmospheric pressure
option 3 Custom Range
Accuracy, in air or nitrogen ±2 °C (±3.6 °F) Td (See the figure below)
When the dew point is below 0 ℃, the transmitter outputs the frost point
response time63% [90%]The gas temperature is+20 °CAnd pressure1 bartime
-60 -> -20°C Td (-76 -> -4 °F Td) 5 s
-20 -> -60°C Td (-4 -> -76 °F Td) 45 s
PPMvolume concentration
measurement range(typical) 10 . .. 40000 ppm
Accuracy,+20 °C (+68°F), 1 bar1 ppm + 20%The response time of the reading63% [90%],
operating environment
Tested gas Non corrosive gas
temperature*) -40 . .. +60°C (-40 ... +140 °F)
relative humidity0 . .. 100 %RH
pressure*) 0 . ..50 bar (725 psia)
Sample gas flow rate No impact on measurement accuracy
*)When the gas temperature drops to-40 °C (-40°F)Or the pressure is high enough50 bar(725 psia)When, the supply voltage must be24 . .. 28 VDC.
output
analog output(The corresponding range can be changed) 4 . .. 20 mA (3line), 0 . .. 1 V/5 V,1 . .. 5 V
Current output resolution0.002 mA
Voltage output resolution0.3 mV
Typical temperature coefficient0.005 %output range/°C
digital outputRS-485,non-isolatedVaisalaindustrial protocol
cable joint4-needleM8 (IEC 60947-5-2)
routine
sensorVaisala DRYCAP® 180D
Suggested calibration cycle
Within the given accuracy range2 years
Supply voltage during voltage output12 . .. 28 VDC
Supply voltage during current output18 . .. 28 VDC
power supply current
During normal measurement maximum10mA +load current
In the self diagnosis stage maximum220 mApulse
Load during current output maximum500 Ohm
Load during voltage output minimum10 kOhm
Shell material stainless steel(AISI316L)
Sensor protection Stainless steel sintered filter
mechanical interfaceISO G1/2” orNPT 1/2
Protection levelIP66 (NEMA 4)
Storage temperature range-40 . .. +60°C (-40 ... +140 °F)
weight pipe thread90g (3.2oz) NPTthread100g (3.5oz)
satisfyEMCstandardEN61326-1Electronic devices are used for measurement, control, and laboratory workBusiness environment.