Drying room dew point analyzer, lithium battery production large screen dew point analyzer

NegotiableUpdate on 12/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
The online dew point meter adopts Vaisala DRYCAP #174; The DMT143L dew point transmitter of the sensor is a compact dew point measuring instrument that can be directly installed in a pressure system of 50 bar (725 psi). Vaisala DRYCAP #174; Sensor technology has long-term high-performance performance.
Product Details

DMT143L online dew point meterBriefly describe

The online dew point meter adopts Vaisala(Vaisala)DRYCAP®sensorDMT143LDew point transmitter is a compact dew point measuring instrument that can be directly installed on50 bar(725 psia)In the pressure system. vaisala(Vaisala)DRYCAP®Sensor technology has long-term high-performance performance.

This sensor is resistant to humid environments, thereforeDMT143LThe transmitter can perform very well in industrial applications with water splashes, including pipeline condensation during process system failures or start-up phases. The sensor is also highly resistant to particulate impurities, oil and gas, and most chemical gases, and is not sensitive to changes in flow rate.

DMT143LOnline Dew Point Meterfeature/benefit

1A compact dew point transmitter suitable for small industrial dryer applications

2Vaisala with automatic calibration function(Vaisala)DRYCAP®Sensor Technology

3The calibration cycle is as long as once every two years

4Dew point measurement range-80... +20ºC(-76...+140°F)

5Accuracy ±2 ºC (±3.6 ºF)

6Anti condensation and dew condensation

7Fast response time

8, with Vaisala(Vaisala)Handheld dew point meterDM70compatible

9Tracing back to the sourceNISTCalibration of(Including calibration certificate)

10ThroughRS485User port, convenient and fast maintenance and data transmission

11Excessive dew pointLEDLamp alarm

DMT143LOnline Dew Point Metertechnical indicators

measurement parameters

dew point temperature

measurement range(typical)

-80 . .. +20°C (-76 ... +140 °F) Td

Analog output range

option1 -80 . .. +20°C (-112 ... +68 °F) Td

option2 -80 . .. +20°C (-112 ... +68°F) TdDew point at atmospheric pressure

option 3 Custom Range

Accuracy, in air or nitrogen ±2 °C (±3.6 °F) Td (See the figure below)

When the dew point is below 0 ℃, the transmitter outputs the frost point

response time63% [90%]The gas temperature is+20 °CAnd pressure1 bartime

-60 -> -20°C Td (-76 -> -4 °F Td) 5 s

-20 -> -60°C Td (-4 -> -76 °F Td) 45 s

PPMvolume concentration

measurement range(typical) 10 . .. 40000 ppm

Accuracy,+20 °C (+68°F), 1 bar1 ppm + 20%The response time of the reading63% [90%]

operating environment

Tested gas Non corrosive gas

temperature*) -40 . .. +60°C (-40 ... +140 °F)

relative humidity0 . .. 100 %RH

pressure*) 0 . ..50 bar (725 psia)

Sample gas flow rate No impact on measurement accuracy

*)When the gas temperature drops to-40 °C (-40°F)Or the pressure is high enough50 bar(725 psia)When, the supply voltage must be24 . .. 28 VDC.

output

analog output(The corresponding range can be changed) 4 . .. 20 mA (3line), 0 . .. 1 V/5 V,1 . .. 5 V

Current output resolution0.002 mA

Voltage output resolution0.3 mV

Typical temperature coefficient0.005 %output range/°C

digital outputRS-485,non-isolatedVaisalaindustrial protocol

cable joint4-needleM8 (IEC 60947-5-2)

routine

sensorVaisala DRYCAP® 180D

Suggested calibration cycle

Within the given accuracy range2 years

Supply voltage during voltage output12 . .. 28 VDC

Supply voltage during current output18 . .. 28 VDC

power supply current

During normal measurement maximum10mA +load current

In the self diagnosis stage maximum220 mApulse

Load during current output maximum500 Ohm

Load during voltage output minimum10 kOhm

Shell material stainless steel(AISI316L)

Sensor protection Stainless steel sintered filter

mechanical interfaceISO G1/2 orNPT 1/2

Protection levelIP66 (NEMA 4)

Storage temperature range-40 . .. +60°C (-40 ... +140 °F)

weight pipe thread90g (3.2oz) NPTthread100g (3.5oz)

satisfyEMCstandardEN61326-1Electronic devices are used for measurement, control, and laboratory workBusiness environment.