The Discovery HP-TGA 750 high-pressure thermogravimetric analyzer is suitable for a wider range of gases and can be applied to tests that require ultra-high heating rates, such as gasification reactions.
Product Details
Discovery HP-TGA 750 High Pressure Thermogravimetric Analyzerimage:
Discovery HP-TGA 750 High Pressure Thermogravimetric AnalyzerFeatures and advantages:
High resolution quality readings, with high resolution to meet micro mass change measurement and fast dynamic measurement.
Small sample measurement chamber, with only about 50ml of measurement chamber, is easier to meet high pressure or high vacuum requirements, has low gas consumption, and is easier to replace with gas.
• Corrosion resistance. Can be applied to a wider range of gases.
• High heating and cooling rates. Can be applied to tests that require ultra-high heating rates, such as gasification reactions.
No need for porous insulation material to come into contact with the reaction gas. Ensure the purity of the reaction gas and rapid replacement.
HP-TGA750 adopts a new magnetic levitation weighing technology with international expertise, combining ultra-high sensitivity, stability, and reliability. The internal cavity is small and the measurement range is wide. Its excellent accuracy and fast temperature and pressure control enable optimal experimental data to be obtained while minimizing the consumption of samples, gases, and time.
In addition to conventional TGA applications, due to its magnetic levitation technology, it has great adaptability to all experiments dealing with small samples or toxic/hazardous gases.
The HP TGA750 is equipped with MFCs for quantitative delivery of various test gases. The high-precision pressure control system can control the pressure range from atmospheric pressure to 8Mpa. An ultra fast reactor has a heating and cooling rate of 250K/minute and a temperature of up to 1100 ℃.