Long term stability of DMT132 dew point transmitter (refrigerated dryer) - using HUMICAP; Technology, not affected by compressor oil and the vast majority of chemicals. An economical dew point transmitter suitable for refrigerated dryers.
DMT132 Dew Point Transmitter (Refrigerated Drying Machine) Basic Description
Economical dew point transmitter suitable for refrigerated dryers
- High accuracy measurement of the range of the refrigerated dryer, with an accuracy of up to±1°C((±1.8°F)
- Long term stability–adoptHUMICAP®technology,Not corroded by compressor oil and the vast majority of chemicals.
- Low power demand, operating voltage range10 . .. 28ambush(direct current)
- Use compatible handheld instrumentsDM70orHM70Convenient for functional verification
- optionalLEDAlarm indicator light
DMT132 Dew Point Transmitter (Refrigerated Drying Machine) Parameter Specifications
|
Dew point measurement |
| ? |
Dew point range |
?-20 . .. +50 °C (-4 ... +122 °F) Td |
| ? |
Accuracy, (+20 ° C (+68 ° F)) |
Within the range of -3... 20 ° C (+26.6...+68 ° F) Td, it is ± 1 ° C Within the range of -15... -3 ° C (+5...+26.6 ° F) Td, it is ± 2 ° C |
| |
sensor |
Vaisala HUMICAP ® 180R |
working environment |
| ? |
Working temperature work pressure relative humidity Sample gas flow rate Tested gas |
?-10 . .. + 60 °C (+14 ... +140 °F) 0 . .. 20 bar 0 . .. 100 %RH No impact on measurement accuracy Non corrosive gas |
output |
| ? |
Analog output (expandable) Current output resolution Analog output accuracy +20 °C |
?4...20 mA, 2 lines 0.002 mA ± 0.05% full range |
| ? |
Typical temperature characteristics Data connection port |
± 0.005% full range/° C 4-针 M8( (IEC 60947-5-2)? |
| |
General Parameters |
| |
Suggested calibration cycle (Used for freezer dryer applications) Mechanical installation interface working voltage external load |
2 years
G 1/2' ISO 10 . ... 28 VDC Maximum 100 ohm, supply voltage<20 VDC The maximum voltage is 500 ohms, with a supply voltage of 20... 28 VDC |
| ? |
weight housing material
Shell protection level
Storage temperature range Startup time
|
65 grams (2.3 ounces) PPS+40% GF (reinforced plastic) IP65 (NEMA 4) -40 . .. + 80 °C (-40 ... +176 °F) 3 s |
| |
Options and attachments |
| |
Tubular filter Special protection kit for HMK15 NPT converter Sampling unit |
230602 230914 210662SP DMT242SC, DMT242SC2, DSC74, DSC74B, DSC74C, DMCOIL |
|