The D07-23 mass flow controller/flowmeter is based on the principle of capillary heat transfer temperature difference calorimetry, with a simulated range set, which is not inaccurate due to temperature and pressure fluctuations, and can achieve accurate measurement and control of various gases without temperature and pressure compensation. The D07-23 series products are high flow analog control products developed and manufactured by Huacheng Electronics. They have the characteristics of high accuracy, easy maintenance, and high stability, and support zero point automatic calibration function.
D07-23 Mass Flow Controller/Flow MeterAccording to the principle of capillary heat transfer temperature difference calorimetry, the simulation range is set to achieve accurate measurement and control of various gases without temperature and pressure compensation, without losing accuracy due to temperature and pressure fluctuations. The D07-23 series products are high flow analog control products developed and manufactured by Huacheng Electronics. They have the characteristics of high accuracy, easy maintenance, and high stability, and support zero point automatic calibration function.
D07-23 Mass Flow Controller/Flow MeterAdvantage:
High flow type: up to 250SLM N2
High precision, high stability, fast response: accuracy less than ± 1% F.S., linearity less than ± 0.5% F.S., repeatability less than ± 0.2% F.S., response speed not exceeding 2s
Low zero drift coefficient: Temperature compensation function is implemented using software and hardware, zero point: ≤± 0.05% F.S./℃; Adjustment: ≤± 0.1% F.S./℃
Fully functional: client software with automatic reset, alarm, soft start, delay and other functions
International certification: obtained CE, ROHS, UL certification
VCP's valve control technology: The valve control technology adopts multiple new achievements, collectively referred to as VCP technology. This structure has less dead space inside the valve, fast valve response speed, and no hysteresis phenomenon. The use of VCP technology can effectively improve the flow stability of gas mass flow controllers
Innovation point:
Circuit board digital core: digital core computing, achieving better control and regulation, and real-time data interaction
Constant power sensor: Adopting a new type of constant power sensor technology, it greatly improves the measurement accuracy and response speed of the system
Digital sensor compensation technology: using digital sensor bridge balance method to compensate for sensor driving voltage and suppress temperature drift
Application fields:
Clean Energy - Photovoltaic
Clean Energy - Hydrogen Fuel Cells
Special Industry - Fiber Optic
Special Industry - Vacuum Coating
Special Industry - Heat Treatment
Life Sciences - Environmental Protection
Life Sciences - Food Industry
Life Sciences - Bioengineering
Life Sciences - Medical and Health
Life Sciences - Analytical Instruments
Scientific research - experimental equipment
Technical Indicators:
| model |
D07-23J type |
| Flow specifications |
20,30,50,100,150,200,250 SLM |
| Flow control range |
2~100% of F.S. |
| accuracy |
±1% F.S. |
| linear |
±0.5% F.S. |
| repeatability accuracy |
±0.2% F.S. |
| response time |
≤ 2sec |
| Range of working pressure difference |
0.05~0.3MPa 20 SLM≤F. S.≤30SLM |
| 0.1~0.3MPa 50 SLM≤F. S.<150 SLM |
| 0.15~0.3MPa 150 SLM≤F. S.≤250 SLM |
| Maximum work pressure |
1 MPa (145psig) |
| pressure resistance |
3 MPa (435.1 psig) |
| leak rate |
1×10-9Pa·m3/sec He |
| temperature coefficient |
Zero: ≤±0.05% F.S./℃; Span: ≤±0.1% F.S./℃; |
| Working environment temperature |
5℃ ~ 45℃ |
| input signal |
4~20 mA or 1~5 VDC |
| output signal |
4~20 mA or 1~5 VDC |
| power supply |
24VDC 400 mA (voltage input range: ± 10% 24V) |
| weight |
2.6 kg |