D07-11 mass flow controller/flowmeter is based on the principle of capillary heat transfer temperature difference calorimetry, with a simulated range set, which is not inaccurate due to temperature and pressure fluctuations, and can achieve accurate measurement and control of various gases without temperature and pressure compensation. The D07-11 series products are high-voltage products developed by Huacheng Electronics, with a maximum pressure resistance of 10MPa.
D07-11 Mass Flow Controller/Flow MeterAccording to the principle of capillary heat transfer temperature difference calorimetry, the simulation range is set to achieve accurate measurement and control of various gases without temperature and pressure compensation, without losing accuracy due to temperature and pressure fluctuations.D07-11 Mass Flow Controller/Flow MeterThe series of products are high-voltage products developed by Huacheng Electronics, with a maximum pressure resistance of 10MPa.
advantage
High pressure resistance: The maximum working pressure can reach 10Mpa
Stable control, high accuracy, good repeatability: accuracy less than ± 1% F.S., linearity less than ± 0.5% F.S., repeatability less than ± 0.2% F.S.
Easy to maintain: The design is simple and easy to use, easy to maintain, and can reduce the cost of use and maintenance.
innovation points
VCP's valve control technology: The valve control technology adopts multiple new achievements, collectively referred to as VCP technology. This structure has less dead space inside the valve, fast valve response speed, and no hysteresis phenomenon. The use of VCP technology can effectively improve the flow stability of gas mass flow controllers.
Constant power sensor: It adopts a new type of constant power sensor technology, which greatly improves the measurement accuracy and response speed of the system.
Application field
Special Industries - Petrochemical Industry
Special Industry - Fiber Optic
Special Industry - Vacuum Coating
Special Industry - Heat Treatment
Life Sciences - Environmental Protection
Life Sciences - Food Industry
Life Sciences - Bioengineering Life
Science - Medical Health
Life Sciences - Analytical Instruments
Scientific research - experimental equipment
technical indicators
| model |
D07-11C |
| Flow specifications |
( 0~5,10,20,30,50,100,200,300,500) SCCM |
| ( 0~1,2,3,5,10,20,30) SLM |
| Flow control range |
2~100% of F.S. |
| accuracy |
±1 % F.S (20,30SLM: ±2 % F.S.) |
| Linear |
±0.5 % F.S (20,30SLM: ±2 % F.S.) |
| repeatability accuracy |
±0.2 %F.S. |
| response time |
≤ 10 sec |
| Range of working pressure difference |
≤10SLM (0.2 ~ 0.8) MPa (29~116psid) |
| >10SLM (0.2 ~ 0.6) MPa (29~87psid) |
| Maximum work pressure |
10 MPa (1450psig) |
| pressure resistance |
10 MPa (1450psig) |
| Working environment temperature |
5 ℃ ~ 45 ℃ |
| input signal |
0 V ~ +5.00 V |
| (Input impedance greater than 100K, output current not greater than 3mA) |
| output signal |
0 V ~ +5.00 V |
| (Input impedance greater than 100K, output current not greater than 3mA) |
| power supply |
+15 V 50 Ma; -15V 200 mA |
| Weight |
1.2kg |