Overview
The BM-RHT-6AS temperature and humidity transmitter is a low operating voltage current output transmitter with a nominal operating voltage of 24V. It is a sensor with two channels (4~20) mA output for temperature and humidity. Made with high-precision calibration free digital output temperature and humidity sensors and low-power microcontrollers, it has short response time, high accuracy, and good stability.
1. Low operating voltage, analog current output
BM-RHT-6AS has a nominal operating voltage of 24VDC and (4~20) mA output, suitable for providing 24V DC field equipment
2. High precision, high stability, high reliability
The BM-RHT-6AS series adopts high-precision calibration free digital temperature and humidity sensors, which have high accuracy, short response time, high stability, and high reliability
3. High load capacity
The BM-RHT-6AS series temperature and humidity transmitter can drive a maximum load resistance of 450 Ω at a working voltage of 24V
4. Low self heating effect
The BM-RHT-6AS series features low operating voltage and thermal isolation design, greatly reducing the impact of transmitter self heating on measurement accuracy.
5. Reverse power protection
The BM-RHT-6AS series temperature and humidity transmitter has a reverse power protection function to prevent equipment damage caused by reverse power polarity.
6. Surge and ESD protection
The BM-RHT-6AS series temperature and humidity transmitter provides 500A surge protection and contact discharge 6kV air discharge 8kV ESD protection, making the product suitable for harsh electromagnetic environments.
7. Application scope:
Computer room, data center, telecommunications cabinet, HVAC, greenhouse, weather station, automation, warehouse, incubation room.
technical indicators
| parameter | condition | minimum | typical value | maximum |
unit |
| Working voltage Vs | RL=0Ω | 15 | 24 | 30 | V |
| load resistor | Vs=24V | 450 |
Ω | ||
| measurement range | Temperature measurement range | -20 | 50 | °C | |
| humidity | 0 | 100 | %RH | ||
| Output current range | Vs=(15~30)V | 4 | 20 | mA | |
| precision | Temperature, @ 10°C~50°C @-20°C~80°C @-40°C~100°C |
0.3 0.5 0.7 |
0.4 0.8 1.2 |
°C | |
| Humidity, @ 10%RH~90%RH 0%RH~100%RH |
3 4.5 |
5 8 |
%RH | ||
| Self heating effect | @Ta=25°C,Ha=75%RH @-20°C~50°C |
0.5 |
0.6 |
°C |
|
| response time | RH(63%) | 8 | s | ||
| Temp(63%) | 15 | s | |||
| working environment | Temperature: -25°C~85°C | ||||
| Humidity: 0%RH~95%RH(No condensation) | |||||
| Overall dimensions | 85mm×85mm×27mm | ||||
| Recommended working environment | Vs=24,RL=250Ω | ||||
Instructions:
(1) The maximum load resistance is calculated according to the formula RL ≤ (Vs-15)/0.02, where Vs is the operating voltage in V and RL is in Ω
(2) The self heating effect is the difference between the temperature value measured within 1 minute of power on and the temperature value measured after 30 minutes of power on. The self heating effect has been corrected for a working voltage of 24V, a load resistance of 250 ohms, an ambient temperature of 25, and a relative humidity of 75%.
(3) Response time refers to the time it takes to apply a step to reach 63% of the step value.
For more product information, please call/