- Phone
-
Address
No. 452 Dongmen Street, Jintan City, Jiangsu Province
Changzhou Tianrui Instrument Co., Ltd
No. 452 Dongmen Street, Jintan City, Jiangsu Province
Digital constant temperature double hole water bath potIt is mainly used for distillation, drying, concentration, and temperature preservation of chemical or biological products in laboratories. It can also be used for constant temperature heating and other temperature tests. It is a tool for biology, genetics, viruses, aquaculture, environmental protection, medicine, health, laboratories, analysis rooms, education and scientific research.
| Model TypeNo | Temperature Control Range (℃) | Power (W) | Working Size (cm) |
| HH-2 two hole | Adjust freely within the range of room temperature+5 ℃ -100 ℃ | 600w | 31*16*12 |
Digital constant temperature double hole water bath potPerformance:
1. Power supply 220V ± 10% 50HZ ± 2%
2. Heating power 600w
3. Temperature control range can be adjusted freely within the range of room temperature+5 ℃ -100 ℃
4. Large error ≤ ± 1 ℃
5. Temperature fluctuation ≤ ± 0.5 ℃
6. Volume size 310 * 160 * 120mm
7. Customizable according to customer requirements
constant temperature water bathUsage and maintenance:
(1) Before use, an appropriate amount of clean tap water must be added to the pot, or hot water can be added at the desired temperature to shorten the heating time.
(2) Connect the power and select the temperature. When equipped with an electronic thermostat, adjust the "temperature knob" clockwise to the desired temperature, which is the heating state and the green indicator light is on. When heated to the desired temperature, the red indicator light will turn on, indicating a constant temperature state. When equipped with a digital display head, the large digit of the counter is ten, and pressing the operation symbol is an increase, and pressing the operation symbol is a decrease. Traffic lights switch with changes in body temperature. Similarly, the green light indicates that the heater is working, while the red light indicates constant temperature. The use of this instrument requires heating and constant temperature at least twice to achieve the correct temperature accuracy (it must be fully covered and sealed).
(3) After completing the work, set the temperature control knob and the controller to low values and cut off the power.
(4) Do not add too much water to prevent the water from overflowing outside the pot when boiling,
(5) The amount of water in the pot should not expose the heating tube to the water surface. To avoid burning out, causing water and electricity leakage.
(6) When using this product, the three pin socket must be effectively grounded.