- Phone
-
Address
Zhenwu Industrial Park, Jiangdu City, Jiangsu Province
Jiangdu Xuanyu Experimental Machinery Factory
Zhenwu Industrial Park, Jiangdu City, Jiangsu Province
JF-5Combustion performance oxygen index tester for building materials wire and cable
It is applicable to the determination of the combustion performance of plastics, rubber, electrical sleeving, fiber, foamed plastics, film, film, textile and other materials.
Advantages of the YZS-8A fully automatic oxygen index analyzer from Xunyu: 1 Full color touch screen control, simply set the oxygen concentration value on the touch screen, and the program will automatically adjust to the oxygen concentration balance and emit a beep sound prompt, eliminating the trouble of manually adjusting the oxygen concentration; 2. The step proportional valve is used to greatly improve the flow control accuracy. The closed-loop control is used to automatically adjust the oxygen concentration drift program to the target value during the test, which avoids the disadvantage that the traditional oxygen index tester cannot adjust the oxygen concentration during the test, and greatly improves the test accuracy. Jiangdu Xuanyu Experimental Machinery Factory
JF-5Combustion performance oxygen index tester for building materials wire and cable Technical parameters of automatic oxygen index analyzer
1. Adopting imported oxygen sensors, the digital display of oxygen concentration does not require calculation, with higher and more accurate accuracy and a range of 0*
2. Digital resolution: ± 0.1%
3. Measurement accuracy: 0.1 level
4. Touch screen setting program automatically adjusts oxygen concentration
5. One click calibration accuracy
6. One click concentration ratio
7. Automatic alarm sound for stable oxygen concentration
8. Equipped with timing function
9. Can store experimental data
10. Historical data can be queried
11. Can clear historical data
12. Can choose whether to burn 50mm or not
13. Gas source fault prompt
14. Oxygen sensor fault prompt
15. Oxygen and nitrogen misconnection prompt
16. Aging prompt for oxygen sensor
17. Input of standard oxygen concentration
18. The diameter of the combustion cylinder can be set (two commonly used specifications are optional)
19. Flow regulation range: 0-25L/min (0-1500L/h)
20. Quartz glass tube: Choose one of two specifications (inner diameter ≥ 75mm or inner diameter ≥ 85mm)
21. Gas flow velocity inside the combustion cylinder: 40mm ± 2mm/s
22. Overall dimensions of the machine: 650mm × 400 × 830mm
23. Pressure gauge accuracy level 2.5, resolution: 0.01MPa
24. Experimental environment: ambient temperature: room temperature to 40 ℃; Relative humidity: ≤ 70%;
25. Input pressure: 0.25-0.3MPa
26. Working pressure: Nitrogen 0.15-0.20Mpa, Oxygen 0.15-0.20Mpa
27. The sample holder can be used for soft and hard plastics, various building materials, textiles, fire doors, etc
28. Propane (butane) ignition system, the ignition nozzle is made of a metal tube, and the tail end has a nozzle with an inner diameter of Φ 2 ± 1mm, which can be freely bent. Can be inserted into the combustion cylinder to ignite the sample, flame length: 16 ± 4mm, size 5mm-60mm can be freely adjusted,
29. Gas: Industrial nitrogen, oxygen, purity>99%; (The purity of industrial oxygen provided by users/regular gas stations is generally between 99.2% and 99.3%).
30. Power requirements: AC220 (+10%) V, 50HZ
31. High power consumption: 150W
32. Self supporting material sample clamp: can be fixed at the axis position of the combustion cylinder and can vertically clamp the sample
33. Non self-supporting material sample clamp: capable of fixing both vertical edges of the sample to the frame at the same time (optional/applied to soft non self-supporting materials such as textiles)
Design standard for the Xunyu fully automatic oxygen index analyzer YZS-8A: GB/T 2406.2-2009 GB/T 2406.1-2008
Compliant with standards: ASTM D 2863, ISO 4589-2, NES 714 GB/T 5454 GB/T 10707 GB/T 8924-2005 GB/T 16581-1996 NB/SH/T 0815-2010 TB/T 2919-1998 IEC 61144-1992 ISO 15705-2002 ISO 4589-2-1996